Using the Calibration Utility
DT9850 Series modules are calibrated at the factory and should not require calibration for
initial use. We recommend that you check and, if necessary, readjust the calibration of the
analog output circuitry on the DT9850 Series modules every six months using the DT9850
Series Calibration Utility and a precision digital multimeter (DMM).
Note:
Ensure that you installed the DT9850 Series Device Driver prior to using the DT9850
Series Calibration Utility.
Start the DT9850 Series Calibration Utility as follows:
1.
Make sure that one or more DT9850 Series modules are attached to your computer.
2.
Attach a precision DMM to the DT9850 Series module’s outputs.
3.
Click
Start
from the Task Bar, and then select
Programs | Data Translation, Inc |
Calibration | DT9850 Series Calibration Utility.
The DT9850 Calibration utility appears.
4.
Select the module to calibrate from the
Board
drop-down list box.
Once the DT9850 Series Calibration Utility is running, you can calibrate the analog output
circuitry of the DT9850 Series module, described on
.
Содержание DT9850 Series
Страница 1: ...DT9850 Series UM 22600 M User s Manual Title Page ...
Страница 4: ......
Страница 8: ...Contents 8 ...
Страница 11: ...11 1 Overview DT9850 Series Hardware Features 12 Supported Software 14 Getting Started Procedure 15 ...
Страница 16: ...Chapter 1 16 ...
Страница 17: ...Part 1 Getting Started ...
Страница 18: ......
Страница 48: ...Chapter 4 48 ...
Страница 49: ...Part 2 Using Your Module ...
Страница 50: ......
Страница 51: ...51 5 Principles of Operation Analog Output Features 53 Digital I O Features 60 Counter Timer Features 62 ...
Страница 76: ...Chapter 6 76 ...
Страница 77: ...77 7 Troubleshooting General Checklist 78 Technical Support 80 If Your Module Needs Factory Service 81 ...
Страница 82: ...Chapter 7 82 ...
Страница 83: ...83 8 Calibration Using the Calibration Utility 84 Calibrating the Analog Output Subsystem 85 ...
Страница 86: ...Chapter 8 86 ...