Chapter 5
74
•
Analog threshold trigger
– For the DT9837 module, the start trigger event occurs when
the signal attached to analog input channel 0 rises above 1.0 V (the fixed threshold level).
Using software, specify the start trigger source as a positive threshold trigger, and the
threshold trigger channel as channel 0.
For the DT9837A and DT9837B modules, the start trigger event occurs when the signal
attached to analog input channel 0 rises above a user-specified threshold value. Using
software, specify the start trigger source as a positive threshold trigger, the threshold
trigger channel as channel 0, and the threshold level as a value between 0.2 V and 9.8 V.
For the DT9837C module, the start trigger event occurs when the signal attached to a
specified analog input channel in the channel list rises above or falls below a
user-specified threshold value. Using software, specify the following parameters:
−
Start trigger source – Specify a positive (low-to-high transition) threshold trigger if you
want to trigger when the signal rises above a threshold level, or a negative
(high-to-low transition) threshold trigger if you want to trigger when the signal falls
below a threshold level.
−
Threshold channel – Specify any one of the analog input channels that are included in
the channel list as the threshold input channel.
−
Threshold level – Specify a value between ±10 V for a gain of 1 or ±1 V for a gain of 10
as the threshold level.
Note:
On the DT9837A, DT9837B, and DT9837C modules, if you configure the
synchronization mode as slave, the RJ45 connector accepts trigger and clock signals from
the master; you cannot use the analog threshold trigger on the slave module. Refer to
for more information.
Reference Trigger Sources
DT9837 Series modules support an analog threshold trigger for the reference trigger.
The reference trigger event occurs when the signal attached to a specified analog input
channel in the channel list rises above or falls below a user-specified threshold value. Using
software, specify the following parameters:
• Reference trigger source – Specify a positive (low-to-high transition) threshold trigger if
you want to trigger when the signal rises above a threshold level, or a negative
(high-to-low transition) threshold trigger if you want to trigger when the signal falls
below a threshold level.
• Threshold channel – Specify any one of the analog input channels that are included in the
channel list as the threshold input channel.
• Threshold level – Specify a value between ±10 V for a gain of 1 or ±1 V for a gain of 10 as
the threshold level.
Содержание DT9837
Страница 1: ...DT9837 Series UM 22417 AJ User s Manual Title Page ...
Страница 4: ......
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Hardware Features 14 Supported Software 21 Supported Accessories 24 Getting Started Procedure 25 ...
Страница 26: ...Chapter 1 26 ...
Страница 27: ...Part 1 Getting Started ...
Страница 28: ......
Страница 49: ...49 4 Verifying the Operation of a Module Select the Device 51 Acquire Data from an Analog Output Channel 54 ...
Страница 53: ...Verifying the Operation of a Module 53 ...
Страница 60: ...Chapter 4 60 ...
Страница 61: ...Part 2 Using Your Module ...
Страница 62: ......
Страница 90: ...Chapter 5 90 ...
Страница 104: ...Chapter 6 104 ...
Страница 105: ...105 7 Troubleshooting General Checklist 106 Technical Support 108 If Your Module Needs Factory Service 109 ...
Страница 110: ...Chapter 7 110 ...
Страница 130: ...Appendix A 130 ...
Страница 144: ...Appendix C 144 ...
Страница 150: ...Index 150 ...