Principles of Operation
73
Figure 23: Continuous Scan Mode on the DT9837 Series Modules
Using a Start and Reference Trigger
Input Triggers
A trigger is an event that occurs based on a specified set of conditions.
On the DT9837 Series modules, you can specify a start trigger source and a reference trigger
source. Pre-trigger data acquisition starts when the start trigger event occurs. When the
reference trigger event occurs, pre-trigger data acquisition stops and post-trigger acquisition
starts. Post-trigger acquisition stops when the number of samples you specify for the
post-trigger scan count has been reached. Refer to
for more information.
Start Trigger Sources
The DT9837 Series modules support the following sources for the start trigger:
•
Software trigger
– A software trigger event occurs when you start the analog input
operation (the computer issues a write to the module to begin conversions). Using
software, specify the start trigger source as a software trigger.
•
External digital (TTL) trigger
– An external digital (TTL) trigger event occurs when the
module detects a rising-edge transition on the signal connected to the Ext Trig connector
on the module. Using software, specify the start trigger source as an external, positive
digital (TTL) trigger.
Note:
On the DT9837A, DT9837B, and DT9837C modules, if you configure the
synchronization mode as slave, the RJ45 connector accepts trigger and clock signals from
the master; you cannot use the Ext Trig connector on the slave module. Refer to
for more information.
Chan 0
Chan 1
Chan 2
Input
Sample
Clock
Chan 0
Chan 1
Chan 2
Chan 0
Chan 1
Chan 2
Chan 0
Chan 1
Chan 2
Start Trigger occurs
Reference Trigger occurs
3 Post-trigger samples acquired
Pre-trigger data acquired
Chan 0
Chan 1
Chan 2
Post-Trigger Scan Count = 3
Chan 0
Chan 1
Chan 2
Trigger Sample is not counted
as post-trigger sample
Содержание DT9837
Страница 1: ...DT9837 Series UM 22417 AJ User s Manual Title Page ...
Страница 4: ......
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Hardware Features 14 Supported Software 21 Supported Accessories 24 Getting Started Procedure 25 ...
Страница 26: ...Chapter 1 26 ...
Страница 27: ...Part 1 Getting Started ...
Страница 28: ......
Страница 49: ...49 4 Verifying the Operation of a Module Select the Device 51 Acquire Data from an Analog Output Channel 54 ...
Страница 53: ...Verifying the Operation of a Module 53 ...
Страница 60: ...Chapter 4 60 ...
Страница 61: ...Part 2 Using Your Module ...
Страница 62: ......
Страница 90: ...Chapter 5 90 ...
Страница 104: ...Chapter 6 104 ...
Страница 105: ...105 7 Troubleshooting General Checklist 106 Technical Support 108 If Your Module Needs Factory Service 109 ...
Страница 110: ...Chapter 7 110 ...
Страница 130: ...Appendix A 130 ...
Страница 144: ...Appendix C 144 ...
Страница 150: ...Index 150 ...