Chapter 8
114
Calibrating the Analog Input Subsystem
This section describes how to use the calibration utility to calibrate the analog input
subsystem of a DT9837 Series module.
DT9837 Series modules have separate calibration for each A/D input channel.
Connecting a Precision Voltage Source
To calibrate the analog input circuitry, you need to connect an ex9.3750 V precision
voltage source to the DT9837 Series module. Connect the precision voltage source to the first
channel you want to calibrate; for example, Analog In 0 (AD Ch0).
Using the Auto-Calibration Procedure
Auto-calibration is the easiest to use and is the recommended calibration method. To
auto-calibrate the analog input subsystem, do the following:
1.
Select the
A/D Calibration
tab of the calibration utility.
2.
Under the
Calibration Settings
area of the window, select the sampling frequency, in
Hertz, and the gain that you want to use.
Note:
By default, this utility uses DC coupling with the current source disabled. This is
the recommended setting for most users. While changing these settings is not advised for
most users, users with unique applications can change these settings by clicking the
Advanced
button.
3.
Under the Automatic Calibration area of the window, select the channel that you want to
calibrate from the
Type of Calibration
drop-down list box, and then click the
Start
button.
A message appears notifying you to verify that 0.000 V is applied to the channel.
4.
Verify that the supplied voltage to your selected channel is 0.000 V, and then click
OK
.
The offset value is calibrated. When the offset calibration is complete, a message appears notifying
you to set the input voltage of the channel to +9.375 V.
5.
Check that the supplied voltage to your selected channel is +9.375 V, and then click
OK
.
The gain value is calibrated and a completion message appears.
6.
Repeat these steps for each analog input channel on the module.
Note:
At any time, you can click
Restore Factory Settings
to reset the A/D calibration
values to their original factory settings. This process will undo any auto or manual calibration
settings.
Содержание DT9837
Страница 1: ...DT9837 Series UM 22417 AJ User s Manual Title Page ...
Страница 4: ......
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Hardware Features 14 Supported Software 21 Supported Accessories 24 Getting Started Procedure 25 ...
Страница 26: ...Chapter 1 26 ...
Страница 27: ...Part 1 Getting Started ...
Страница 28: ......
Страница 49: ...49 4 Verifying the Operation of a Module Select the Device 51 Acquire Data from an Analog Output Channel 54 ...
Страница 53: ...Verifying the Operation of a Module 53 ...
Страница 60: ...Chapter 4 60 ...
Страница 61: ...Part 2 Using Your Module ...
Страница 62: ......
Страница 90: ...Chapter 5 90 ...
Страница 104: ...Chapter 6 104 ...
Страница 105: ...105 7 Troubleshooting General Checklist 106 Technical Support 108 If Your Module Needs Factory Service 109 ...
Страница 110: ...Chapter 7 110 ...
Страница 130: ...Appendix A 130 ...
Страница 144: ...Appendix C 144 ...
Страница 150: ...Index 150 ...