Wiring Signals
53
5
5
5
5
5
5
5
5
5
STP268 Screw Terminal Assignments
Screw terminals on the STP268 screw terminal panel are available for
low-current signal conditioning applications. The supply on the
DT3010-268 board is current-limited through a 10
Ω
resistor and is
specified for a maximum load current of ±3 mA.
The +5.0 V reference on the STP268 screw terminal panel is also
current-limited through a 10
Ω
resistor and is provided for
applications that require a reference less than 1 mA.
The +5 V output signal on the STP268 screw terminal panel is
current-limited through a series 10
Ω
resistor and supports loads up
to 100 mA. Note that you must take the drop (current [I] multiplied
by resistance [R]) across the series 10
Ω
resistor (1 V at 100 mA) into
consideration.
Note:
If you are connecting a high-speed clock to the STP268, it is
recommended that you connect the return to the adjacent ground
screw terminal.
lists the screw terminal assignments on the STP268 screw
terminal panel for connector J1 of the DT3010-268;
lists the
screw terminal assignments on the STP268 screw terminal panel for
connector J2 of the DT3010-268.
Содержание DT3010 Series
Страница 1: ...R DT3010 Series UM 16868 C Getting Started Manual ...
Страница 11: ...1 1 Overview DT3010 Series Key Hardware Features 2 DT3010 Series Software 3 Getting Started Procedure 4 ...
Страница 22: ...Chapter 2 12 ...
Страница 36: ...Chapter 3 26 ...
Страница 114: ...Chapter 5 104 ...
Страница 136: ...Appendix A 126 ...
Страница 142: ...Index 132 ...