Chapter 6
112
5.
In the Channel list box, select analog output channel 0.
6.
In the Range list box, select the output range of DAC0.
The default is ±10 V.
7.
Enter an output value, or use the slider to select a value to output
from DAC0.
8.
Click Send to output a single value from DAC0.
The value that is output is displayed on the screen in both text and
graphical form.
Performing a Continuous Analog Input Operation
To verify that the board can perform a continuous analog input
operation, perform the following steps:
1.
Connect known voltage sources, such as the outputs of a function
generator, to analog input channels 0 and 1 on the DT3010 Series
board (using the differential configuration). Refer to
for
an example of how to connect a differential analog input.
2.
Click the Acquisition menu.
3.
For this example, click Scope.
4.
Select the appropriate DT3010 Series board from the Board list
box.
5.
In the Sec/Div list box, select the number of seconds per division
(.1 to .00001) for the display.
6.
In the Channels list box, select analog input channel 1, then click
Add
to add the channel to the channel list.
Channel 0 is automatically added to the channel list.
7.
Click Config from the Toolbar.
8.
From the Config menu, select ChannelType, then select
Differential
.
Содержание DT3010 Series