Rerunning the PV-Kit
If you suspect a problem with the
FLX500
, you
can check to see if the problem is the
FLX500
and
not a device, algorithm, or power source, by
rerunning the PV-Kit job. All hardware required
to run the PV-Kit job, including 16 devices, came
with the
FLX500
. Contact Data I/O for informa-
tion about ordering replacement parts for the
PV-Kit.
NOTE: Be sure the Socket Adapters are
clean before running the PV-Kit job.
1.
Rerun the PV-Kit job. For instructions see
2.
Read the
Throughput
displayed on the sta-
tistics screen on the monitor by tapping the
Statistics
button
on the Run Window.
Record the PPH in the PV-Kit Performance
chart on page 47.
Figure 14—The Statistics Window displays
Throughput, often abbreviated as PPH.
If the PPH is the same as when the machine was
new (approximately 500 PPH) then you can
eliminate many of the possible error causes.
Содержание FLX500
Страница 4: ... Data I O FLX500 ii ...
Страница 12: ... Data I O FLX500 Chptr 1 8 blank page ...
Страница 28: ... Data I O FLX500 Chptr 2 24 blank page ...
Страница 40: ... Data I O FLX500 Chptr 4 36 blank page ...
Страница 42: ... Data I O FLX500 Chptr 5 38 Daily Sign off Clean the sockets page 10 Date Initials Notes ...
Страница 43: ...5 Sign off Charts Daily Sign off 39 Daily Sign off Date Initials Notes ...
Страница 44: ... Data I O FLX500 Chptr 5 40 Daily Sign off Date Initials Notes ...
Страница 45: ...5 Sign off Charts Daily Sign off 41 Daily Sign off Date Initials Notes ...
Страница 46: ... Data I O FLX500 Chptr 5 42 Daily Sign off Date Initials Notes ...
Страница 48: ... Data I O FLX500 Chptr 5 44 Weekly Sign off Date Initials Notes ...
Страница 52: ... Data I O FLX500 Chptr 5 48 blank page ...
Страница 57: ... FLX500 ...
Страница 58: ... FLX500 FLX programmable media experts ...