EASY
SENSE Vu+ Reference Manual DO255 (3)
25
Data Harvest Group Ltd.
Fitting the lanyard
Hard Reset
If Vu+ fails to respond to the computer or a button press, carry out a Hard Reset.
1. If necessary attach Vu+ to USB power.
2. Press
+
+
at the same time.
If the hard reset has been done correctly, the LCD will display
Vu+
before returning to the main menu.
Note:
A Hard Reset does
not
erase any stored data sets but if a recording is interrupted by a reset, then the data
being captured will not be stored.
If Vu+ still fails to respond apply a Full Reset (see below) or contact Technical Support at Data
Harvest.
Please provide details of:
The computer platform it is being used with.
The software version number
A description of the problem being encountered
If possible, telephone from a location where you can operate Vu+ with the computer.
Full Reset
A Full Reset will return Vu+ back to the factory default.
It will:
Erase
all stored data sets.
Automatically return the built-in sensors to their default range (Light - Indoor 1 k Lux,
Temperature -
0 to 50˚C).
Return the date format to DD-MM-YY.
Return decimals to On with a full stop format.
Return the language to its default.
Scroll
▲▼
the menu to
System
, press enter
►
to select. Scroll down
▼
to
Full Reset
. Press the
enter
►
button to select.
Select enter
►
to confirm this selection.
Select enter
►
to return to the System menu.
Thread the thin cord loop through both
slot holes in the base of the logger.
Thread the ribbon strap completely through
the open loop of the thin cord. Pull the thin
cord end tight to achieve a secure fitting.