WARNINGS & PRECAUTIONS
The RadiSense system is a closed loop fibre system and therefore classified as a Class 1M laser
system according to EN60825-1:2014 and EN60825-2:2005. In order to provide laser safety in
case of a fibre failure or accidental disconnection of the fibres, the RadiSense system is provided
with an Automatic Laser Shutdown (ALS) as described in the EN60825-2:2005.
Although the RadiSense system does not emit any laser light under normal
circumstances, it is good common practice never to look into any of the fibre
optic connectors.
As a safety precaution, products that use a LASER can only be turned on using
a LASER Code. This code can be entered using the touchscreen of the
RadiCentre system.
(In combination with the RadiCentre CTR1004B or CTR1009B)
As a safety precaution, products that use a LASER can only be turned on by
pressing the ‘START’ button of the RadiCentre system for several seconds. An
auditory warning will sound during this activation period.
(In combination with the RadiCentre CTR1001S)
Make sure that the fibre cables are installed correctly before activating the
system. Do not activate the system if the fibre optic cables show any sign of
damage or tampering.
To make the RadiSense
as safe as possible, the plug-in card and sensors have
their own safety interlock system that is designed to work with the RadiCentre.