The use of the mounting angle is optional. For isotropic measurements of an electrical field of
which the polarization is unknown (for example in reverb chamber / mode-stir chambers) the
mounting angle is not needed, see left picture.
For accurate measurements of electrical field with a known polarization, the mounting angle can
be used to position one axis of the RadiSense parallel to the field. See right picture.
In most situations, this is the way an E-field sensor is calibrated, axis by axis. Therefore, the probe
should be used in the same way to obtain the most accurate results.
REMARK:
Please be noted that placing dielectric material in close proximity to the field sensor will influence
the field and could cause measurement errors.