DARAY V409 Скачать руководство пользователя страница 9

NOTE 

 

Make sure to place the device on the finger in a correct direction. The LED part of the 
sensor should he at the backside of the patient hand and the photodetector part at 
the inside. Make sure to insert the finger to a suitable depth into the sensor so that 
the fingernail is just opposite to the light emitted from the sensor. 

 

To acquire accurate results, please read data until the device is steadily placed. 

 
Measurement Limitations 
If the accuracy of any measurement does not seem reasonable, first check the patient's vital 
signs by an alternate method. Then check the instrument for proper function. 

Inaccurate measurements may be caused by: 
 

 

High-frequency electrical noise, such as noise from electrosurgical apparatus connected 
to the system; 

 

Significant levels of dysfunctional hemoglobins (e.g., carboxyhemoglobin or 

methemoglobin); 

 

Significant concentrations of dysfunctional hemoglobin, such as carboxyhemoglobin and 
methemoglobin; itravascular dyes such as indocyanine green or methylene blue; 

 

Exposure to excessive illumination, such as surgical lamps (especially ones with a xenon 
light source), bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight 
(exposure to excessive illumination can be corrected by covering the device with a dark 
material); 

 

Excessive patient motion; 

 

Venous pulsations; 

 

SpO2 is too low; 

 

Improper position of the finger inserted into the device; 

 

Placement of the device on the same extremity with a blood pressure cuff, arterial 
catheter, or intravascular line. 

 

Loss of pulse signal can occur in the following situation: 

 

The device clip is too tight; 

 

There is excessive illumination from light sources such as a surgical lamp, a bilirubin 
lamp, or sunlight; 

 

A blood pressure cuff is inflated on the same extremity as the one with the device 

attached: 

 

The patient has hypotension, severe vasoconstriction, severe anaemia. or hypothermia; 

 

There is arterial occlusion proximal to the device; 

 

The patient is in cardiac arrest or in shock. 

  
Precautions 

WARNING 

 

Check if the device is in normal condition before monitoring. Do not use the device 

once it is found damage. 

 

Remove the device from the patient after measurement. 

 

Do not put the device on extremities with arterial catheter or venous syringe. 

 

The device can not be clipped on the oedema and tender tissue. 

 

If no pulse is found or the reading is unreasonable, first check the patient 's condition, 
and then check the device placement on the finger, finally consult a qualified 
engineer to check the device for correct function. 

 

Don't use the device to measure patients whose pulse rate is lower than 30bpm, 
which may cause incorrect results. 

 

The device is not appropriate for the ceaseless monitoring of the patients. Prolonged 
and continuous monitoring may increase jeopardy of unexpected change of dermal 

condition such as abnormal sensitivity, ruhescence, vesicle, repressive putrescence, 
and so on. The uncomfortable or painful feeling may appear if using the device 

Содержание V409

Страница 1: ...QAM V409 0616 4 V409 Fingertip Pulse Oximeter...

Страница 2: ......

Страница 3: ...ODUCTION 6 4 APPEARANCE AND DISPLAY 7 5 INSTALLATION 7 6 OPERATION 7 7 SpO2 MEASUREMENT 8 8 MAINTENACE 9 9 DISPOSAL 11 10 TROUBLESHOOTING 11 11 SPECIFICATIONS 12 Returns policy 13 Warranty 14 WARRANTY...

Страница 4: ...ides application tips or other useful information to ensure that you get the most from your product WARNING The device is intended only as an adjunct in patient assessment It must be used in conjuncti...

Страница 5: ...ry power It is compact tiny light low power consumption and easy for learning and handling It is suitable for monitoring adult and paediatric patients It is widely used in the hospital s operation roo...

Страница 6: ...Displays SpO2 pulse rate pulse bar graph and waveform 3 SpO2 waveform 4 Pulse bar graph Has 6 levels Indicates the pulse strength of SpO2 measurement 5 Clip press symbol Press here to open clip 6 Low...

Страница 7: ...ls as shown in figure 5 2 Cover the lid onto the cabinet and push it upwards to make it close well Figure 5 1 Battery Cabinet Figure 5 2 Installing Batteries CAUTION Use AAA alkaline batteries or rech...

Страница 8: ...r side The sensor measurement wavelengths are nominally 660nm for the Red LED and 940nm for Infrared LED Maximum optical power output for LED is 4 mW The amount of light transmitted depends on many fa...

Страница 9: ...ed into the device Placement of the device on the same extremity with a blood pressure cuff arterial catheter or intravascular line Loss of pulse signal can occur in the following situation The device...

Страница 10: ...the device can not be powered on or when powering on the battery capacity bar is 0 level or the low battery power indicator is red and flashes replace the batteries with good ones In case of any damag...

Страница 11: ...olluted by chemical substance the users should handle it effectively according to the properties of the chemical substance WARNING The cleaning solutions above can only be used for general cleaning If...

Страница 12: ...ettle the patient The device can not be turned on 1 The batteries are drained or almost drained 2 The batteries are not installed correctly 3 There is a problem with the device 1 Change batteries 2 Re...

Страница 13: ...cedure OTHER Any situation which is not covered by the above We will try to help but we cannot normally offer a refund DOA dead on arrival For additional clarification please refer to our terms and co...

Страница 14: ...aray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from c...

Страница 15: ......

Страница 16: ......

Страница 17: ......

Отзывы: