6 Operation 

Press the power button to turn on the device. The OED screen appears display. If there is no 

finger inserted into the device, the device will automatically shut down in 8 seconds to save 
power. 
 

NOTE 

The device is powered by batteries. The batteries must be installed in the device before 

powering on. If the battery power is insufficient, the device may fail to be turned on. At 
this moment, the batteries should be replaced with new ones. 
 

7. SpO2 Measurement 

Measurement principle 
SpO2 plethysmogram measurement is employed to determine the oxygen saturation of 
hemoglobin in the arterial blood. The SpO2/PLETH parameter can also provide a pulse rate 
signal, pulse strength and a plethysmogram wave. How the SpO2 I PLETH Parameter Works 

 

 

Arterial oxygen saturation is measured by a method called pulse oximetry. It is a 
continuous, non-invasive method based on the different spectra absorption of hemoglobin 
and oxyhemogiobin (called spectrophotometer principle). It measures how much light, 
sent from light sources on one side of the sensor, is transmitted through patient tissue 

(such as a finger or a toe), to a receiver on the other side. 

 

The sensor measurement wavelengths are nominally 660nm for the Red 

 

LED and 940nm for Infrared LED.  
Maximum optical power output for LED is 4 mW. 

 

The amount of light transmitted depends on many factors, most of which are constant. 
However, one of these factors, the blood flow in the arteries, varies with time, because it 
is pulsating. By measuring the light absorption during a pulsation, it is possible to derive 
the oxygen saturation of the arterial blood. Detecting the pulsation gives a PLETH 

waveform, pulse rate signal and pulse strength. 

 

The SpO2 value, PR value and pulse strength can be displayed on the screen. 

 

SpO2 is a non-invasive measurement of the functional oxygen saturation. 

 

The light sources and the receiver of the sensor are laid in the two rubber cushions of the 

device respectively. 

  
Measurement steps 

 

Press (5) in the Figure3-1 and open the clip. 

 

Insert the patient's finger into the rubber cushions of the clip, make sure the finger is in 
the right position as shown in figure 6-1, and then clip the finger. 

 

Press the power button on the front panel to turn on the device. 

 

Using first finger, middle finger or ring finger when doing test. 

 

Do not shake the finger and keep the patient at ease during the process. 

 

The readings will be displayed on the LED screen momentarily. 

  

 

Figure 6-1 Put Finger Into the correct position 

Содержание V409

Страница 1: ...QAM V409 0616 4 V409 Fingertip Pulse Oximeter...

Страница 2: ......

Страница 3: ...ODUCTION 6 4 APPEARANCE AND DISPLAY 7 5 INSTALLATION 7 6 OPERATION 7 7 SpO2 MEASUREMENT 8 8 MAINTENACE 9 9 DISPOSAL 11 10 TROUBLESHOOTING 11 11 SPECIFICATIONS 12 Returns policy 13 Warranty 14 WARRANTY...

Страница 4: ...ides application tips or other useful information to ensure that you get the most from your product WARNING The device is intended only as an adjunct in patient assessment It must be used in conjuncti...

Страница 5: ...ry power It is compact tiny light low power consumption and easy for learning and handling It is suitable for monitoring adult and paediatric patients It is widely used in the hospital s operation roo...

Страница 6: ...Displays SpO2 pulse rate pulse bar graph and waveform 3 SpO2 waveform 4 Pulse bar graph Has 6 levels Indicates the pulse strength of SpO2 measurement 5 Clip press symbol Press here to open clip 6 Low...

Страница 7: ...ls as shown in figure 5 2 Cover the lid onto the cabinet and push it upwards to make it close well Figure 5 1 Battery Cabinet Figure 5 2 Installing Batteries CAUTION Use AAA alkaline batteries or rech...

Страница 8: ...r side The sensor measurement wavelengths are nominally 660nm for the Red LED and 940nm for Infrared LED Maximum optical power output for LED is 4 mW The amount of light transmitted depends on many fa...

Страница 9: ...ed into the device Placement of the device on the same extremity with a blood pressure cuff arterial catheter or intravascular line Loss of pulse signal can occur in the following situation The device...

Страница 10: ...the device can not be powered on or when powering on the battery capacity bar is 0 level or the low battery power indicator is red and flashes replace the batteries with good ones In case of any damag...

Страница 11: ...olluted by chemical substance the users should handle it effectively according to the properties of the chemical substance WARNING The cleaning solutions above can only be used for general cleaning If...

Страница 12: ...ettle the patient The device can not be turned on 1 The batteries are drained or almost drained 2 The batteries are not installed correctly 3 There is a problem with the device 1 Change batteries 2 Re...

Страница 13: ...cedure OTHER Any situation which is not covered by the above We will try to help but we cannot normally offer a refund DOA dead on arrival For additional clarification please refer to our terms and co...

Страница 14: ...aray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from c...

Страница 15: ......

Страница 16: ......

Страница 17: ......

Отзывы: