9. Disposal 

 
To avoid contaminating or infecting personnel, the environment or other equipment, make 

sure you disinfect or decontaminate the device appropriately before disposing of it in 
accordance with your country's law for equipment containing electrical and electronic parts.  
 

10. Troubleshooting 

Problem Possible 

reason

Solution

The SpO2 and PR are 
not displaying 
normally. 

1. The finger is not properly 
positioned. 
2. The patient's SpO2 is too low to 
be detected. 

I. Place the finger properly and try 
again. 
2. Try again: Get a qualified 
medical diagnosis if you are sure 
the device is functioning correctly. 

SpO2 and PR display 
are unstable. 

1. The finger is not placed inside
enough 
2. The finger is shaking or the 

patient is moving 

I. Place the finger property and try 
again. 
2. Settle the patient. 

The device can not 
be turned on. 

1. The batteries are drained or 
almost drained 

2. The batteries are not installed 
correctly 
3. There is a problem with the 
device. 

1. Change batteries.
2. Reinstall batteries correctly. 

3. Contact the supplier. 

 

Please contact the supplier if any other problems are found. 
 
  

11. Specifications 

 

General 

Monitoring parameters: SpO2, PR, Pulse strength/waveform 
Display mode: 64k colour OLED 
Low battery power indicator: OLED 
Size: 63 (H) x 34 (W) x 33 (D) mm 

Weight: 55 (including batteries) 
  

Electrical specifications 

Working voltage: 2.2V DC ~ 3.4 V DC 
Battery type: 2x 1.5V AAA alkaline 
 

Environment 

Temperature 
Operation: 0 ~ 40 (C) 

Transportation and storage: -20 ~ 60 (C) 
 
Humidity (noncondensing) 
Operation: 15% ~ 95 %  
Transportation and storage: 10% ~ 95 % 
 

Altitude 
Operation: 86 ~ 106 KPa 
Transportation and storage: 50 ~ 106 KPa 

 

Parameter Specifications 

SpO2 
Patient: Adult, paediatric 
Range: 35 ~ 99% 
Resolution: 1% 
Accuracy: ±2% (90%~00%), ±3 % (70%~9%) 
 

PR 
Range: 0 ~ 250bpm 
Resolution: 1bpm 
Accuracy: ±2bpm

 

Содержание V409

Страница 1: ...QAM V409 0616 4 V409 Fingertip Pulse Oximeter...

Страница 2: ......

Страница 3: ...ODUCTION 6 4 APPEARANCE AND DISPLAY 7 5 INSTALLATION 7 6 OPERATION 7 7 SpO2 MEASUREMENT 8 8 MAINTENACE 9 9 DISPOSAL 11 10 TROUBLESHOOTING 11 11 SPECIFICATIONS 12 Returns policy 13 Warranty 14 WARRANTY...

Страница 4: ...ides application tips or other useful information to ensure that you get the most from your product WARNING The device is intended only as an adjunct in patient assessment It must be used in conjuncti...

Страница 5: ...ry power It is compact tiny light low power consumption and easy for learning and handling It is suitable for monitoring adult and paediatric patients It is widely used in the hospital s operation roo...

Страница 6: ...Displays SpO2 pulse rate pulse bar graph and waveform 3 SpO2 waveform 4 Pulse bar graph Has 6 levels Indicates the pulse strength of SpO2 measurement 5 Clip press symbol Press here to open clip 6 Low...

Страница 7: ...ls as shown in figure 5 2 Cover the lid onto the cabinet and push it upwards to make it close well Figure 5 1 Battery Cabinet Figure 5 2 Installing Batteries CAUTION Use AAA alkaline batteries or rech...

Страница 8: ...r side The sensor measurement wavelengths are nominally 660nm for the Red LED and 940nm for Infrared LED Maximum optical power output for LED is 4 mW The amount of light transmitted depends on many fa...

Страница 9: ...ed into the device Placement of the device on the same extremity with a blood pressure cuff arterial catheter or intravascular line Loss of pulse signal can occur in the following situation The device...

Страница 10: ...the device can not be powered on or when powering on the battery capacity bar is 0 level or the low battery power indicator is red and flashes replace the batteries with good ones In case of any damag...

Страница 11: ...olluted by chemical substance the users should handle it effectively according to the properties of the chemical substance WARNING The cleaning solutions above can only be used for general cleaning If...

Страница 12: ...ettle the patient The device can not be turned on 1 The batteries are drained or almost drained 2 The batteries are not installed correctly 3 There is a problem with the device 1 Change batteries 2 Re...

Страница 13: ...cedure OTHER Any situation which is not covered by the above We will try to help but we cannot normally offer a refund DOA dead on arrival For additional clarification please refer to our terms and co...

Страница 14: ...aray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from c...

Страница 15: ......

Страница 16: ......

Страница 17: ......

Отзывы: