User Guide | Gas detecting sensor, type PAGM for CO2 and Halocarbon

© Danfoss | DCS (vt) | 2021.03

4 | BC372016561395en-000101

Note:

 When the Carbon Purge/Zero Air filter is saturated and can no longer absorb refrigerant, 

refrigerant can pass through and offset the baseline measurements of the unit which will than result in 

unit going into fault condition <0100> error will appear. Please adhere to filter maintenance intervals 

as below recommended.

2.3 Battery

3.  Front Panel Display & 

Controls

The monitor is powered by a rechargeable battery pack located in a pouch on the side of the monitor’s 

soft carrying case. The battery pack can be recharged at any time, regardless of the battery’s current 

charge state.  The lithium-ion battery will likely be partially charged. However, you should fully charge 

the battery prior to operating the PGM-IR. Refer to the battery manufacturer’s instruction sheet for the 

proper charging procedure.

Before the unit can be used please connect the battery pack the unit is designed for a 16 V DC input. 

Set battery voltage output to 16 V (refer to the battery manufacturer’s instruction sheet for details).

A fully charged battery pack will power the monitor for a minimum of 8 hours.

ALARM (Red LED)

DISPLAY SCREEN

ENTER key
1.  Press to zero the peak reading
2.  When in the Control/Setup menu 

mode, press to save a displayed 
value

KEYPAD
When in the Control/Setup menu mode, 
use these buttons to:
1.  Move the arrow (>) on the display 

screen to the desired function

2.  Scroll through data
3.  Change a function’s value

Enter the Control/Setup menu mode 
by pressing the left and right arrow 
buttons simultaneously

POWER ON/OFF (toggle switch)

Carrying handle

SYSTEM FAULT (Yellow LED)

MONITOR ON (Green LED)

(flashed during warm-up)

SILENCE/ESC

Press this button to:

1. Temporarily silence the audible alarm

2. Return to the previous screen without 

saving data

RS-232 connector

Tubing connection

Coalescing Inline Filter Element

Power
switch

Purge/Fresh Air
Carbon Filter

Gas

sample

port

Zero air 

inlet port

Exhaust 

port

Battery pack

connector

Содержание PAGM

Страница 1: ...User Guide Gas detecting sensor Type PAGM for CO2 and Halocarbon www danfoss com...

Страница 2: ...xhaust and zero air filter assembly 3 2 1 Gas Sample Probe 3 2 2 Purge Zero Air Carbon Filter Assembly 3 2 3 Battery 4 3 Front Panel Display Controls 4 4 General operation 5 4 1 Normal operation 5 4 2...

Страница 3: ...tions For details on appropriate use refer to the general description application and operation discussions in this manual Before risking equipment damage or personal injury contact the manufacturer i...

Страница 4: ...Before the unit can be used please connect the battery pack the unit is designed for a 16 V DC input Set battery voltage output to 16 V refer to the battery manufacturer s instruction sheet for detail...

Страница 5: ...ng functions MEASURE indicates that the monitor is actively measuring the refrigerant gas level at the sample point To reset the peak value to zero press the ENTER button PURGE is displayed when the m...

Страница 6: ...ng the SILENCE button while a fault condition still exists causes the internal audible alarm to turn OFF for a period of time as set by the SILENCE The front panel FAULT light will continue to flash O...

Страница 7: ...act service department for further instructions 0800 Sample Flow Fault Check for a restriction in the gas sample inlet or exhaust Check the external Coalescing Inline Filter and or internal Hydrophobi...

Страница 8: ...ied out at least once every 6 months but is dependent on the environment and usage If used in harsh environment where a large amount of dust and moisture particles are present in atmospheres the maint...

Страница 9: ...ient CO2 level typically 390 ppm Example unit exposed to 5000 ppm should read 4610 ppm 5 which is from 4380 4840 ppm PAGM IR for halogen gas Set up the unit to measure R134A gas Using a regulator infl...

Страница 10: ...Danfoss DCS vt 2021 03 BC372016561395en 000101 10 ADAP KOOL...