DANA Diabecare RS Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Quick guide

 

IFUQ-120-EN (rev.1_180109)

 

Содержание Diabecare RS

Страница 1: ...1 Quick guide IFUQ 120 EN rev 1_180109 ...

Страница 2: ...2 This page is left blank intentionally ...

Страница 3: ...he Quick Reference Guide QRG is supplementary to the IFU Instruction For Use All precautions and warnings within the IFU must be read and adhered to always Read the entire IFU before commencing or using the DANA Diabecare RS Insulin Pump Refer to the DANA RS IFU for advanced features instruction and more complex operation of the DANA Diabecare RS Insulin Pump ...

Страница 4: ... is correctly inserted the pump turns on automatically and performs self checks to make sure all the systems are working properly 1 Use the battery key to open the battery cap by turning counter clockwise 2 Insert the battery with terminals facing down 3 Tighten the battery cap using the battery key until it is firm and flush with the top of the pump When changing the battery clean the inside thre...

Страница 5: ...new UI menu works in a circular rotation through the menu s using 1 Press to increase insulin dosages time and date 2 Press to decrease insulin dosages time and date or to exit to previous menu cancel stop button 3 Press to move to the next function on the LCD screen or to scroll to the right in menus 4 Press to select the function highlighted on the LCD screen Confirmation button to confirm setti...

Страница 6: ... displayed on the DANA RS home screen Date Time Current basal rate in units per hour The flashing indicates the basal is active Percentage of basal rate 100 is normal basal delivery without a temporary rate active Battery Indicator Units remaining in the pump reservoir ...

Страница 7: ...7 Pump Menu ICON s ...

Страница 8: ...ed and engaged 4 Push the linking screw and barrel of the reservoir to the top ready to fill with insulin This process ensures that the double O ring is moving freely and will not cause an occlusion within the pump 5 Remove the needle protective cap and insert into the 3ml vial of room temperature insulin 6 Using the needle protective cap push the insulin plunger down so that the insulin is slowly...

Страница 9: ... and then select the prime option from within the refill menu and attach the new Infusion set tube to the pump Start the prime 14 Stop or suspend the prime when the tube is full of insulin and there are no air bubbles visible in the tube Using the key will pause the prime 5cm of insulin tube 1 unit insulin 15 If using a Teflon plastic cannula select the prime cannula menu this fills the empty need...

Страница 10: ...Basal profile Adjust the start time and end time for the section period being changed or amended Note TDD Total Daily Dose Next adjust the Basal rate u hr for the period selected When press to save a SETTING SAVED screen appears Press to finish the setting or press to move to the start time to set next Basal rate To confirm changes select ...

Страница 11: ...cific to individual preferences such as pump alerts as beep or vibration and clock display as 12 or 24 hour format Note it is necessary to scroll through all options even if no changes are made and confirm and all changes by selected Exit and confirming YES to save options configured When setting or changing any of the selected options the or key changes the selected option The key moves to the ne...

Страница 12: ... On Time that LCD backlight remains on in between button press 7 Language Default language options for pump operation and menu 8 Glucose Unit mmol or Mgdl usually default setting based upon country of origin 9 Shutdown Time setting before pump alarms due to no button presses If alarm not acknowledged basal will suspend 10 Low Reservoir Personalized low volume level for alarm 10 20 30 units in 10u ...

Страница 13: ...tion menu Select the first option CIR CF from setting menu Set and personalize CF and CIR ratios for each specific time of day A Healthcare Professional and or Insulin Pump trainer will configure some additional settings within the DANA Dr Mode These include daily hourly and bolus maximums Basal and Bolus increments decreasing ratio duration of active insulin and target or Ideal BG ...

Страница 14: ... a standard food bolus without including a correction dose and without reducing the dose for Active Insulin If enabled extended bolus in user options the pump will offer a range of bolus types 1 Step is a standard bolus 2 Extended enables the bolus to be delivered over a long period 3 Dual Pattern is a combination of both above bolus types Select to start the delivery ...

Страница 15: ...d Ideal BG Carb to Insulin ratio and Correction factor Each of these can be altered or select OK to confirm settings The pump calculates the suggested dose G C A This can be adjusted up or down if change is needed Select to start the delivery DANA Insulin Pumps are set to display Blood Glucose in either mmol L or mg dl depending upon the region they are sold some regions the option can be changed ...

Страница 16: ...ended temporary rate and the percentage of standard basal rate 150 is a 50 increase in standard rate 80 is a 20 reduction in standard rate Select to start the TEMP RATE Stopping a temporary rate Select TEMPORARY BASAL from the main menu The TEMP STATE screen is displayed press to STOP the TEMPORARY RATE Confirm the STOP TEMP RATE with ...

Страница 17: ...n Pump 7 Ensure to test BG levelsfrequently especially during long haul flights 8 On all flights carry non liquid hypo treatment in accessible carry on luggage 9 Get comprehensive travel insurance and ensure that it covers diabetes and wearing an Insulin Pump and make sure it is named listed on the policy 10 Change the pump time to destination time once at cruising altitude Complete this for each ...

Страница 18: ...ssivation film which separates the chemicals within the battery This prevents any loss of energy whilst in storage This passivation film gets stronger the longer from manufacture When first loading a new battery the DANA pump startup procedure draws a high current across the battery to purposefully break the passivation film Unfortunately sometimes this just doesn t complete the process perfectly ...