5
Set the cell operating voltage
Operating voltage Usens
Type of
Cell
Min.
Ideal
Each type of cell has a maximum operating voltage value
not to exceed to prevent cell from quickly becoming
unusable.
Max.
DAL17 1.925
2.00
Man-CanDetectors-Cal-En-V10.doc
Can Bus Gas Detectors -
C lib
ti
2.025
DAL50 2.400
2.50
Referring to this table, always adjust the operating voltage
on the "Ideal" value without exceeding the "Max" value.
2.550
VQ41 1.900
2.00
2.050
The type of cell is specified on the marking label on the
detection head of the detector.
e
For the following procedure, the sensor cell as well as the power supply must be connected to the DEX CAN
transmitter. Please refer to the note of installation of the DEX CAN for more information.
Figure 11 : Measuring cell operating voltage on DEX CAN transmitter
•
Power on the DEX CAN transmitter and
wait for at least 15 sec.
•
Set the voltmeter on Vdc position and
connect it between terminals
A
and
P
of
the DEX CAN transmitter.
Note: the transmitter terminal block has special
sockets to receive the voltmeter probes or test
plugs with 2.3mm diameter (Dalemans art. code
BOR00000075).
Figure 12 : Measuring cell operating voltage on remote detector
•
If the DEX CAN transmitter is remote from
the detector, connect the voltmeter
between terminals
A
and
P
of the detector
terminal block.
Figure 13 : Adjusting the cell operating voltage
•
With the precision screwdriver gently
adjust the "
Usens
" potentiometer of the
transmitter so that the voltmeter reading
matches the recommended operating
voltage for the type of cell (refer to column
"Ideal" in table here above).
Voltmeter
+ -
Special sockets
for probes
Voltmeter
+ -
Remote detector
DEX CAN Transmitter
OctoBUS 64 central unit
Usens
potentiometer
Warning !
Potentiometers
are not multi-turn.
Act SOFTLY.