Dakewe(Shenzhen)Medical Equipment Co,. Ltd
41
V4.1
It is strictly forbidden to use an external heater to dry the freezer, otherwise it will cause
damage to the refrigeration system.
7.6 Replacing Component
7.6.1 Replacing the Fuse
When you need to replace the fuse, open the back cover of the machine, pull out the electric control
box, open the upper cover of the electric control box, take out the damaged fuse from the drive
board, and insert a new fuse of the same specification into the fuse holder , And then restore the
machine to its original condition.
1. Power supply cable socket.
2. Compressor overload protection.
Precautions:
Before replacing the fuse, the main switch of the machine must be switched off and the
power plug must be unplugged.
The insurance specification is 4A, the size is 5*20mm, only the fuse of the same
specification can be used. Otherwise, it may cause serious damage to the equipment.
7.6.2 Replacing the UV Lamp
There is a light tube in the freezing room, and the ultraviolet germicidal lamp tube. When
replacing the lamp, the lamp can be rotated 90° and then pulled out from the lamp holder. The
new tube of the same specification can be installed in the reverse order.
Precautions:
Before replacing the UV lamp, the main switch of the machine must be switched off and
the power plug must be unplugged.
If the UV lamp is broken, it must be replaced by a professional.
Figure 29 The picture of
UV lamp ballast