2.2.3 Warning - Operation
2.2.4 Warning - Cleaning and Maintenance
2.3 Equipment Placement Requirements
Ground must be flat.
No vibration on the ground.
There are three-hole power sockets near the instrument placement position.
No obstacles near the hand wheel.
The ambient temperature range is 5 °C ~ 30 °C, when the ambient temperature
reaches 26 °C or above, the air conditioner should be open for cooling.
Air humidity does not exceed 60%.
No sunlight for placing.
Be careful when handling the slicing knife, the blade is extremely sharp, and it is very
serious once being injured! It is highly recommended to wear safety gloves.
Be sure to remove the blade before removing the knife holder from the instrument. When
the blade is not in use, it should be put back into the knife box.
No matter where the blade is placed, the blade must never be upward. Never touching the
falling blade!
Be sure to clamp the sample block first, then clamp the blade.
The handlebar must be locked and the blade must be guarded with the blade guard during
operation or replacement of the blade and sample and during the rest periods!
Be sure to wear safety goggles when making fragile sample sections! Sample may be
fragmented.
Make sure that liquid does not enter the inside of the instrument during operation!
Immediately picking up and removing the embedding agent that has fallen on the floor. It
can cause slipping and there is a risk of injury!
Before slicing, check that the cartridge is firmly clamped on the sample - failure to follow
this instruction may result in damage of the sample.
Before each maintenance or cleaning, the instrument shall be TURNED OFF and the main
power supply shall be disconnected.
Only authorized qualified personnel can access the internal components of the instrument
for maintenance and repair!
⚫
Do not touch falling accessories – there is a risk of injury!
⚫
Lock the hand wheel before cleaning!
⚫
Do not use any solvents containing acetone or xylene for cleaning!
⚫
Ensure that liquid does not enter the instrument during the cleaning process!