UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
Problem
Probable cause
Solution
Ultrasound
-The equipment doesn’t work.
- Chair’s fuse burned.
- Turn off the chair from mains
power and request a Technician
presence.
- L a c k o f p o w e r t o t h e
ultrasound.
- Deformed insert.
- Loosen insert.
- Bad utilization (incorrect
attack angle).
- Change the insert.
- Hold the insert with the key
- See item “Technical and
applications”.
- There is no water in the hand
piece.
- Inadequate alimentacion
pressure water.
- Bad regulating of the water
flux.
- Correct the water filter.
- Adjust the water flux through
the actuator.
Bicarbonate Jet
- Insufficient bicarbonate
in the jet.
- Lack of bicarbonate in the
recipient.
- Obstruction in the exit of the
recipient or in the beak.
- Excess of bicarbonate in the
recipient.
- Inadequate position of the jet.
- Add bicarbonate into the
recipient (maximun. of 40g).
- Remove the blocked parts
with a plunger.
- Take out the exceed.
- S e e “ Te chni cal and
Applications”.
- Lack of pressure in the jet.
- Compressor off.
- Turn on the compressor.
- Lack of water in the jet.
-Registration of closed water.
-Jet Water Reservoir empty.
-Open the record.
-Get the reservoir water.