CAUTION
Do not turn on the air supply before all connections have been
made and checked.
NOTICE
The system is not fitted with lifting eye bolts or other equipment for
lifting with telphers or similar equipment. The system should only
be lifted using the lower part of the frame.
CAUTION
The product is designed for indoor use only.
System operation
WARNING
High pressure. The flow rate may under no circumstances exceed
the specified column maximum flow rate. High flows might affect
the packed resin, causing the pressure to exceed the specified
column maximum pressure.
WARNING
Operating limits. Never exceed the operating limits stated in this
document and on the product label.
WARNING
Power failure of connected systems. During a power failure of a
connected system, or if the
EMERGENCY STOP button is pressed
on connected equipment, the equipment may remain pressurized.
Opening a line or vessel at this point could result in the release of
potentially hazardous process or cleaning fluid, and cause bodily
harm.
When recovering from a power failure or emergency shutdown of a
connected system, make sure all lines and vessels are depressur-
ized before opening.
2 Safety instructions
2.1 Safety precautions
Chromaflow Packing Station 50, 100, 200, and 400 Operating Instructions 29046228 AD
13