67
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Electrical Utilities
The excel V+ system is available in several electrical configurations. Electrical power should be set up
according to the model ordered. The two line wires in the conduit shall be connected to the building
power (both have live voltages), and the green/yellow wire must be connected to ground.
Systems Designed for Use Outside of Europe
The 120 VAC configurations must be supplied from a dedicated 120±10% VAC, single-phase, 50/60
Hertz source. The wiring should be rated for 20 amps (as per local codes) “Hospital Grade” (NEMA L-
20P or NEMA L-30P). The system can be installed with a removable or lockable wall plug. The 120 VAC
configuration may be changed to a 220 VAC configuration. This internal change must be made by a
Cutera-certified service representative. Call your Cutera representative for information.
The 220 VAC configuration must be supplied from a dedicated 220±10% VAC, single-phase, 50/60 Hertz
source. The wiring should be rated for 16/20 Amps (as per local codes) “Hospital Grade” (NEMA L6-20P
or NEMA L6-30P). The system can be installed with a lockable wall plug. Such a connection will ensure
compliance with allowable leakage current levels per IEC 60601-1 for this device. Leakage current for
these systems does not exceed 500
A.
Systems Designed for Use in European Communities Under MDD
In order to comply with the European Communities Medical Devices Directive 93/42/EEC and
harmonized standard EN 60601-2-22, the 220 VAC configured excel V+ system must be either
permanently connected to a 16A, 220 VAC 50 Hz supply mains in accordance with national wiring
regulations, or connected by means of a dedicated single phase, 16A, 220V wall socket and lockable
plug combination designed to ensure that the connection is “mechanically secured against accidental
loosening”. Such a connection will ensure compliance with allowable leakage current levels per IEC
60601-1 for this device. Leakage current for these systems does not exceed 500
A.
Removable or Lockable Wall Socket and Plug Configurations
If your system is installed with a removable plug or wall socket and lockable plug configuration prior to
installation, the customer’s engineer or electrical contractor will be responsible for ensuring that the
proper electrical requirements are available at the site.
Systems Designed for Use in European Communities Under MDD
To comply with the European Communities Medical Devices Directive 93/42/EEC, the 16A, 250V wall
socket and lockable plug combination must comply with EN 60309 (=IEC 309).
The wall socket and plug combination must meet the following specifications:
Plug:
MK LN 9024, 16A 220V, IP67, IEC 309
Wall Socket:
MK LN 9324, 16A 220V, IP67, IEC 309
Содержание Excel V+
Страница 1: ...I Operator Manual...
Страница 8: ...viii EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Страница 14: ...6 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 Remote Interlock Plug Footswitch...
Страница 104: ...A 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Страница 110: ...C 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Страница 112: ...D 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...