Remove the plastic film from face of the
monitor, and install batteries.
Confirm that the green “IN USE” light is flashing.
Attach the linen clip to the patient’s clothing
(out of the patient’s reach)
Adjust the volume control in the battery
compartment, close the cover, and
tighten the screw
TEST THE ALARM: Remove the magnetic
dome from the socket. The alarm will sound,
and the red “ALARM” light will flash.
Adjust the string to a length that will pull out
the magnetic dome to activate the alarm if the
patient attempts an exit.
CONNECTING TO NURSE CALL: Plug a Nurse
Call Connection Cable into the ”NURSE CALL”
socket on the monitor, connect the opposite
end of cable into Nurse Call Station (an adaptor
may be required) and test.
Attach the Velcro strap.
Secure the monitor to back of the patient’s
chair or bed.
1
1
4
2
2
5
6
3
3
Setup
©2015 Curbell Medical Products, Inc. MAP1212B
www.curbellmedical.com •
1-888-222-2188
CSM-PM100 QUICK SETUP GUIDE
NOTE:
Consult the “Directions For Use” included with monitor for
complete instructions, troubleshooting, and safety information
Use