Cycle Times
39
Cat. No. 01024000
Cycle Times
Use the deionizer controller to set the following cycle times for cations and anions: backwash, slow rinse, fast rinse, quali-
ty rinse, and quality delay. The default settings should be used for most installations; make changes only if necessary.
NOTE If the system uses two deionizer tanks, the user-specified cycle times apply to both tanks.
SERVICE
FEB-01-13 12:11P
10.0GPM 161PPM
1. From the default home screen, press
to display the main menu.
>1)INFORMATION
2)MANUAL REGEN
3)SET DATE/TIME
4)ADV. SETUP
2. From the main menu screen, press
to 4) ADV. SETUP.
3. Press
. The screen displays the Advanced Setup menu.
1)SYSTEM SETUP
2)REGEN SETUP
>3)CYCLE TIMES
4)AUXOUT SETUP
4. Press
and
to select CYCLE TIMES.
BACKWASH CAT
>10
MINUTES
5. The first cycle times setting determines the duration of the cation backwash cycle.
Press
and
or
to change the setting (0–99 minutes).
6. Press
. The screen displays the cation slow rinse setting.
SLOW RINSE CAT
>60
MINUTES
7. Change the acid draw times and other cycle times settings as needed (see below).
Cycle Times Settings
See
"Regeneration Time Cycles" on page 63
Screen Display
Range
Description/Changing the Setting
BACKWASH CAT
>10
MINUTES
0–99 Minutes
(default=10)
The first regeneration cycle is the Backwash Cation Cycle. Press
and
or
then
to change the setting.
ACID DRAW
>30
MINUTES
0–99 Minutes
(default=30)
The duration that the acid pump is on. Press
and
or
then
for setting information.
SLOW RINSE CAT
>45
MINUTES
0–99 Minutes
(default=45)
The duration of the cation slow rinse. Press
and
or
then
to change the setting.