Electrical Installation
CAUTION! Observe the precautions listed below before beginning the electrical installation of your
deionizer controller. Failure to do so may cause permanent damage to the controller.
•
Follow the local electrical code requirements.
•
Be sure electrical power is off and disconnected at the source before completing any wiring/cabling connections.
•
Provide a dedicated 120 Volt circuit for the Smart Controller system to ensure maximum electrical protection.
•
DO NOT include the Smart Controller wiring cables in any conduit or raceway containing other 120 Volt or higher
circuits.
•
Maintain a distance of at least 10 feet between the Smart Controller and any electrical distribution panels, race-
ways carrying 300 Volts or more, and electrical motors of 1 horsepower or more.
•
Use the cabling provided. Failure to do so may effect performance of the Smart Controller adversely.