for conformance to IEC 60601-1 safety standards. When used with other equipment, the
leakage currents may be additive. The person creating the ME system is responsible to comply
with the applicable safety regulations and safety standards for their location. Connect only
IEC 60601-1 approved equipment to this device. Never touch peripheral equipment
connections on this device and the patient at the same time. This can create a risk of shock to
the patient.
•
Do not use the equipment in the prescence of flammable liquids, gases, or other materials
susceptible to ignition due to electrical sparking.
•
This equipment generates and radiates EMI which may affect the normal function of nearby
installed equipment. Equipment that generates and radiates EMI may affect the normal
function of the HD Camera System. Place the equipment away from strong EMI sources.
•
Any serious incident that has occurred in relation to this device should be reported to the manufacturer
and the competent authority of the Member State and/or Country in which the user and/or patient is
established.
Cautions
To prevent improper use and/or damage to this device, please note the following cautions:
•
Carefully unpack this unit and check if any damage occurred during shipment. If damage is
detected, refer to the Warranty and Return Policy section of this manual.
•
This equipment generates heat and uses cooling fans during normal use. To reduce the risk of
overheating and possible loss of function and/or damage to the equipment, install the
equipment in a well-ventilated space within the operating conditions listed herein.
•
There is a risk of damage to the equipment if the equipment is subjected to cleaning or
sterilization methods not approved be the manufacturer. To reduce risk of loss of funtion
and/or damage to the equipment, only use the approved cleaning and sterilization methods
described herein.
•
High frequency surgical devices may cause interference with the video image displayed on
the surgical monitor. To reduce this risk, high frequency devices and imaging equipment
should be located on separate supply circuits.
•
There are no user serviceable parts inside the console or camera heads. There are hazardous
voltages present inside the console. Do not remove the cover. Return the device to the
manufacturer for service.
•
Ensure that the HD Camera System is setup in a manner that any cables that are connected
to the console or head are not a tripping hazard during use.
NOTE:
The warranty will become void if any of these warnings or cautions are disregarded.