38
4.10.4 How to Use the CU-EM1
①
Turn the device on by pressing the Power Button.
When the power of the CU-EM1 is turned ON, the following occurs in sequence:
Beeper: The beeper will beep for 0.5 seconds.
Connection Indicator: The blue indicator will light.
②
Attach the 3-Lead ECG Measurement Cable to the patient.
Attach the 3-lead disposable ECG pads.
③
Turn on the Monitor Mode in the CU-SP2.
For further details on using the Monitor Mode, please refer to [Section 4.8: Monitor Mode] in
these Instructions for Use.
4.10.5 Where to Attach ECG Pads
•
RA/R
: Below the right collarbone
•
LA/L
: Below the left collarbone
•
LL/F
: Left side torso
Using expired disposable ECG electrodes with damaged packaging will not guarantee
accurate measurement of ECG.
The disposable ECG electrodes must be firmly adhered to the patient's skin. Keep the
attachment areas dry.
The conductive parts of the ECG electrodes and associated connectors for applied parts,
including the neutral electrode, should not contact any other conductive parts including earth.
The ECG electrodes are disposable. Do not reuse them.