Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 31 of 81
E.
Press the INCREMENT
▲
button during the five second interval after pressing the
TEMP SET button.
1.
A beeper sounds each time it is pressed, but does not beep when held down
continuously.
2.
The set point display changes; the numbers move up the scale. The highest
setting is 42°C/107°F.
F.
To complete this test routine, press the Power switch to the “O” position.
1.
The control panel goes blank.
2.
The water flow indicator stops.
G.
Disconnect the power cord, loosely coil it and attach it to the nylon straps on the
back of the unit.
H.
Check the blanket/ pad for leaks. Water leaks present a risk of infection to the
patient because the water is not sterile. If a leak is found, the blanket/ pad should
not be used.
I.
Check the couplings at the unit and at the blanket/ pad for positive connection.
J.
To drain the water from the reusable blanket/pad simply shut off power to the
unit and allow the water to drain from the blanket/pad back into the NORM-O-
TEMP
®
unit.
For reusable PLASTIPAD blanket/pad(s), disconnect the connecting hose from
the unit. Loosely coil the hose lengthwise in the center of the blanket/ pad. Fold
the blanket/ pad lengthwise into the center, 1/3 from the left side and 1/3 from
the right side.
For reusable Gelli-Roll blankets/pads either lay them flat or roll the blanket/pad.
Do not fold the Gelli-Roll blankets/pads as folding may damage blanket/pad.
For single-use blanket/pad(s) follow the instructions packaged with the
blanket/pad.
The NORM-O-TEMP
®
unit, connecting hose, and blankets/pads are now ready for
patient use.
3-3.
Unit and Patient Related Precautions
This unit requires both distilled water and electricity to operate.
NOTE:
Please consult pages 4-6 at the beginning of the manual for a complete list of
warnings and cautions related to the NORM-O-TEMP
®
.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...