Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 30 of 81
3-2.2.
Completing a System Test Routine
After arranging the equipment described in Section 3-2.1, complete this System Test
Routine which describes what switches to press and the changes to observe.
ALARM NOTE: If power is removed from the unit without actuating the power switch
(I/O switch) a power fail alarm is activated and the LED located on the front panel of
the unit flashes. To stop alarm, restore power to the unit and push the SILENCE ALARM
button.
ALARM NOTE: If at any time the water falls below the minimum level, the LO WATER
LED is activated and the alarm sounds. The unit shuts down and the operator cannot
proceed until this is corrected by filling the reservoir.
A.
Make sure that the power switch is in the “I” position (unit is turned on).
1.
A beeper sounds.
2.
The microprocessor board goes through self-test.
3.
The seven segment display reads CAL for several seconds.
4.
All segments of the seven segment display flash on and off (8888’s) for a few
seconds along with the LED’s.
5.
A beeper sounds.
6.
The seven segment display flashes the set point temperature 42°C for a few
seconds and the unit begins to pump and heat the water.
7.
The seven segment display will show the actual water temperature.
8.
The HEAT LED will light up if the water temperature is less than 42°C/107.6°F.
9.
The water flow indicator is spinning (only occurs if blanket/pad is attached
properly).
B.
Press C/F button.
1.
A beeper sounds.
2.
The seven segment display changes from 42°C to 107°F.
C.
Press the TEMP SET button.
1.
A beeper sounds.
2.
The seven segment display will begin to flash the set point temperature
42°C/107°F. Each time the operator presses the TEMP SET button after just
having turned on the unit, the set point temperature will display 42°C/107°F.
3.
After pressing the TEMP SET button, the operator has five seconds to change
the set point temperature by pressing either the INCREMENT
▲
or
DECREMENT
▼
buttons.
D.
Press the DECREMENT
▼
button during the five second interval after pressing the
TEMP SET button.
1.
A beeper sounds each time it is pressed but does not beep when held down
continuously.
2.
The set point display changes; the numbers move down the scale. The lowest
set point is 20°C/68°F. However, the operating range of the device is from
32°C/89.6°F to 42°C/107.6°F.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...