Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 11 of 81
NORM-O-TEMP
®
Operating Instructions Quick Start Guide
Read Operation Manual before using this device
First Time Use Cleaning:
Refer to Section 5-1.1, 5-1.2, and 5-1.3 for cleaning instructions.
Initial Operation:
1. Check reservoir water level and fill if necessary.
2. Insert power cord plug into hospital grade outlet.
3. Connect hyperthermia blanket or pad.
4. Turn power switch “ON”.
Unit will go through an 8-10 second calibration cycle to test all indicators and will flash
the set point temperature of 42°C. Observe Water Flow Indicator on the side of the unit to
verify that the pump is circulating water when blanket/pad is connected. The unit will
start heating to a preset temperature of 42°C/107°F. To change the desired set point
temperature, proceed to step 5 below.
5. Depress the”TEMP SET” button. The display will flash the previous set point temperature
or the default temperature of 42°C/107°F if the unit has been turned OFF, and then back
ON. NOTE: You have five seconds to depress one of the arrows or the temperature
readout will return to the actual reservoir water temperature.
6. Depress the “UP”
▲
or “DOWN”
▼
arrow to raise or lower the set point temperature to
the desired setting.
7. Select the desired temperature scale using the “C/F” button (Celsius or Fahrenheit).
Draining Instructions:
1. Turn “OFF” power switch.
2. Connect drain hose to one of the outlet fittings and allow to gravity drain.
Alarm Conditions:
•
If an alarm sounds, check indicator lights or display for problem and take corrective
action.
•
Alarm may be silenced temporarily for five minutes by depressing the “Silence Alarm”
button
•
The “Silence Alarm” button will not silence the alarm when “ERR” appears on the
display or “HIGH TEMP” indicator light comes on and does not flash. Power cord
should be removed from outlet and the unit sent to Biomedical Engineering for repair.
CAUTION
:
•
This device should be operated only by trained personnel, under the direction of a
physician.
•
Patients vary in degree of sensitivity to cold, heat, and pressure
•
The patient’s temperature and skin conditions should be checked at least every 20
minutes, or as directed by a physician, while on a thermal blanket.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...