52/60
Fig 49 - Position - 0,05
Fig 50 - Po 0,05
5.4.2
FUNCTIONING CHECK WITH MEASURING KNOB ON 2
CAUTION
This is the most important checking procedure. The ocular pressure
in this area is highly significant. This check should be carried out at
least twice a year.
1
Position the functioning check accessory in position 2 (first
lateral marking) and the measuring knob on the corresponding
value 2.
2
Turn the measuring knob on the value + 1,95 (measure
equivalent to the width of the calibration sign on the
measuring knob).
3
The arm should be displaced in stop position facing the
examiner.
4
Turn the measuring knob on the value + 2,05 (measure
equivalent to the width of the calibration sign on the
measuring knob).
5
The arm should be displaced in stop position facing the patient.