101
4.4.2 Smart bump
Select this option to undertake smart bump on all sensors when the gas monitor is inserted into
I-Test.
A smart bump will test that the gas monitor responds correctly to a specified level of test gas within a
given time window dependant upon the sensor response time. These parameters are set in the Inlet
Configuration in I-Test Manager, see
i
Only one type of bump test can be selected, if speedy bump test is selected then smart
bump will be disabled and vice versa.
To select a Smart Bump in the sequence, from the Test Sequence screen use the arrow keys to
navigate to the Smart Bump icon
then press
OK
to highlight the Smart Bump icon
À
.
A box will now appear around the icon indicating that Smart Bump function has been selected and
this test will be performed if a gas monitor is inserted into the I-Test.
X
Press
to return to Test Settings screen and then press
again to return to main menu
screen.
Note: Smart bump functionality is only applicable to Gas-Pro, a T4 I-Test test screen will not display
the smart bump icon
.
Figure 82: Smart Bump screen
13 May 2013 11:30:45
I-Test
À
Contents
Exit
Unpacking
Set-up
Operating I-Test
I-Test
Manager
Configuring
I-Test
Warranty
Overview
Contacts
Appendices
Troubleshooting
Specification