Figure 2.29: The TDC calibration tool as provided with the Ndigo device driver.
5. When the
Write Calibration Data
button becomes enabled press it to update the calibra-
tion data on the card.
6. Calibration done!
The card can only be successfully calibrated if:
First Bin
is in the range of 4 to 16
Empty Bins
is less than (First Bin + 4)
at least 10,000 events have been captured
a valid serial number is set.
Important note:
If the application reports an error check if the input pulse is within speci-
fication.
cronologic GmbH & Co. KG
24
Ndigo5G User Guide