Instruction Manual

Instruction Manual

Instruction Manual

Instruction Manual

Instruction Manual

Instruction Manual

Instruction Manual

Instruction Manual

UPPER ARM

BLOOD PRESSURE MONITOR

Model BPM701

This manual and product are not meant to be a substitute for advice provided by 
your doctor or other medical professional. You are not to use the information 
contained herein, or this product for diagnosing or treating a health problem or 
prescribing any medication. If you have or suspect that you have a medical 
problem, promptly consult your healthcare provider.

The device is designed for non-invasive measurements and monitoring of adults’ 
arterial blood pressure. It measures human's systolic blood pressure, diastolic blood 
pressure and heart rate on the arm by oscillometric measurement method.
The device is recommended for use by people over the age of 18 and is for home 
use and not clinical use.

Blood Pressure Monitor Intended Use

*Attention!

To operate the device properly and get the best result from it, please read the 
Instruction Manual thoroughly before you begin measuring and keep it carefully 
so that you can access and consult it at any time.

Medical Disclaimer

1-1. What is blood pressure ?

1. About Blood Pressure

Having one's blood pressure measured by a doctor in a hospital or a clinic, and a 
group health checks, tends to stimulate nervousness in the subject and may even 
create high blood pressure, blood pressure also varies in accordance with a variety of 
conditions and so judgment is not possible on the basis of a single measurement. 
The blood pressure measured first thing in the morning after getting up, before taking 
any food and with the subject still, is known as the fundamental blood pressure. In 
practice it is rather difficult to record the fundamental blood pressure, but to come as 
near as possible to measuring the blood pressure in an environment that is close to 
this is why it is useful to take the measurement at home.

Blood pressure is a measurement of the force of blood flowing against the walls of 
the arteries. Arterial blood pressure is constantly changing during the course of the 
cardiac cycle. The highest pressure in the cycle is called the systolic blood pressure 
and the lowest is the diastolic blood pressure. Both of pressure readings, the systolic 
and the diastolic, are necessary to enable a physician to evaluate the status of a 
patient's blood pressure. Many factors such as physical activity, anxiety or the time of 
day can influence your blood pressure. Blood pressure is typically low in the morning 
and increases from the afternoon to the evening. It is lower in the summer and higher 
in the winter.

1-2. Why is it useful to measure blood pressure at home?

A. WHO blood pressure classi

cations

Standards for assessment of high or low 
blood pressure without regard to age, have 
been established by the World Health 
Organization (WHO), as shown in the chart.

Optimal

Normal

High normal

Mild hypertension

Moderate hypertension

Severe hypertension

 

 

 

 

 

 

 

 

 

120

130

85
80

95

100

90

105

110

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

140

150

160

170

180

* Figure No.1

B. Variations in blood pressure

* Figure No.2

Individual blood pressures vary greatly 
both on a daily and a seasonal basis. 
These variations are even more pronounced in hypertense patients. 
(Hypertense: means a person who has high blood pressure symptom.) 
Normally the blood pressure rises while at work 
and is at its lowest during sleeping period. The 
graph below illustrated the variations in blood 
pressure over a whole day with measurement 
taken every five minutes. (The thick line 
represents sleep. The rise in blood pressure at 4 
PM <A in the graph> and 12 PM <B in the 
graph> correspond to an attack of pain.)

 Please rest for at least 5 ~ 10 minutes before taking the measurement.

 To allow your blood vessels to return to the condition prior to taking the 

 measurement, please wait at least 3 ~ 5 minutes between measurements. You 
 may need to adjust the wait time according to your personal physiological situation.

 If you have one of the circulatory problems as arteriosclerosis, diabetes, liver 

 disease, kidney disease, severe hypertension, peripheral circulation......, please 
 consult your doctor or healthcare professional before using the devices.

 Blood pressure measurements should be interpreted by a physician or trained 

 health professional who is familiar with one's medical history. By using the unit 
 regularly and recording the results for the physician to interpret, one can keep 
 physician informed of the continuing trends in one's blood pressure.

 Wrap the cuff snugly around your arm, the cuff must be at the same level as your 

 heart.

 Do not move the unit during measurement, or accurate measurement can not be 

 achieved.

 Take measurement in a relaxed position without talking.

2. Important information before use of the unit

 Do not wrap the cuff over jacket / sweater sleeve, or measurement cannot be done.

 Keep in mind that blood pressure naturally varies from time to time throughout the 

 day and is affected by lots of different factors such as smoking, alcohol 
 consumption, medication, and physical activities.

 Blood pressure measurements taken with the unit are equivalent to those obtained 

 by a trained observer using thecuff / stethoscope auscultation method and are 
 within the accuracy limits prescribed by the American National Standard for 
 Electronic or Automated Sphygmomanometers.

 Wait 30 ~ 45 minutes before measurement if you have just consumed caffeinated 

 beverages or smoked cigarettes.

When the low battery symbol appears on the display, please change the 
batteries.
Replace all the batteries with new ones and do not mix new and old batteries. 
Do not mix alkaline, standard 
(carbon-zinc) or rechargeable 
(cadmium) batteries either. It may 
shorten the battery life or cause the 
device to malfunction.
Remove the battery cover at the back 
of the unit and insert 4 pieces AA alkaline batteries into battery compartment.
Make sure the polarities “+” and “–“ ends are properly positioned.

4. Inserting/ Changing Batteries

* Figure No.3

3-1. Name of the parts

3. Device Description

3-2. Description of display symbols

* Figure No.4

HOUR: MINUTE

SYSTOLIC PRESSURE

BLOOD PRESSURE UNIT

DIASTOLIC PRESSURE

HEART RATE

PULSE SYMBOL

MONTH/ DATE

LOW BATTERY

MEASURING ERROR SYMBOL

MEMORY NUMBER

*Attention!

 Batteries are hazardous waste. Do not dispose them together with the 

 household garbage. Please take the used batteries to the recycling collection 
 point according to your local regulations.

 Keep the battery away from small children.

 To prolong the battery life and prevent damage caused by leakage, remove 

 the batteries from the device if the device is not to be used for a long period.

 Memories (if any) will not be deleted during battery replacement.

 After replacing the batteries, reset the date and time.

7. Maintenance

7-1. Clean the device body, case and the cuff carefully with a clean, soft and slightly 
 dampened cloth and then wipe them dry with another piece of dry cloth.
7-2. Do not use thinner, alcohol, petrol (gasoline) or any strong cleansers to clean.
7-3. Do not wash the cuff or make it wet, and never use chemical cleaner on it.
7-4. Do not in any way twist the cuff.
7-5. Make sure no water will enter the main unit.
7-6. When the unit is not to be used for a long time, remove the batteries. (Leaking of 
 battery liquid can cause damage.)
7-7. Do not attempt to disassemble or change any parts of the monitor, including cuff.
7-8. If any suggestion or service is requested please consult your service station.
7-9. Do not place the item directly under the sunlight, in high temperature, or humid and 
 dusty places.
7-10. Do not store in extremely low (less than – 20 °C) or high (more than 70 °C) 
 temperature.
7-11. To avoid accidental strangulation, keep this product away from children and do not 
 drape tube around neck.

Model Number
Method of Measurement
Range of measurement

Accuracy

Inflation
Exhaust
Display
Dimension
Unit Weight
Sets of memory
Cuff size
Operating Temperature
Storage Temperature
Power Supply
Battery Life
Auto Power Off
Accessories

BPM701
Oscillometric
Pressure 0 ~ 300 mmHg
Pulse 40 ~ 199 beats/minute
Pr/- 3 mmHg
Pulse +/- 5% Max
Automatic inflation (Air Pump)
Automatic exhaust valve
Liquid Crystal Display
14.5 x 10.6 x 6.5 cm (L x W x H)
Approx. 355 g with batteries
3 databases of 40 memories each = total 120 sets
Arm circumference approx. 23 ~ 33 cm/ 9 ~ 13 inch
+10 °C to + 40 °C, less than 85% R.H.
- 20 °C to + 70 °C, less than 85% R.H.
"AA" (1.5 V) Alkaline battery x 4
About 6 months when used once a day
Whenever not used for 1 minute
4 pcs "AA" (1.5 V) battery, arm cuff with tube, 
instruction manual, pouch.

These specifications are subject to change without notice for purpose of improvement.

9. Speci

cations

0197

This blood pressure monitor complies with the EC Directive and bears the CE 

mark “CE0197”. This blood pressure monitor also complies with mainly 
following standards,(included but not limited) 

Safety standard: 

EN 60601-1 Medical electrical equipment part 1: General requirements for safety

EMC standard: 
EN 60601-1-2 Medical electrical equipment part 1-2: General requirements for safety- Collateral 

 standard: Electromagnetic compatibility- Requirements and tests
Performance standards: 

EN 1060-1 Non-invasive sphygmomanometers - General requirements

EN 1060-3 Non-invasive sphygmomanometers - Supplementary requirements for 
 electromechanical blood pressure measuring systems. 

EN 1060-4 Non-invasive sphygmomanometers - Test procedures to determine the overall 
 system accuracy of automated non-invasive sphygmomanometers.

Important /Note! Read the operating 

instructions.

Consult instructions thoroughly before 

use.

Classification:

- Internally powered equipment

- BF type applied part
- IPX0

- Not suitable for use in presence of 

 flammable anaesthetic mixture with 
 air or with Oxygen or nitrous oxide

- Continuous operation with short-time 
 loading
To avoid inaccurate results caused by 
electromagnetic interference between 

electrical and electronic equipments, do 
not use the device near a cell phone or 

microwave oven.

Discard the used product to the recycling 

collection point according to local 
regulations.

Authorized Representative in the 

European Community

EMERGO EUROPE 

Molenstraat 15, 2513 BH, 

The Hague, The Netherlands
TEL : +31-70-3458570 

FAX : +31-70-3467299

Manufacturer:

HEALTH & LIFE Co., Ltd.
9F, No. 186, Jian Yi Road, 

Chung Ho City 235, Taipei, Taiwan

Distributor:

Sabre Nederland b.v.
Ellermanstraat 23 

1099 BX Amsterdam the Netherlands

te

1. LCD DISPLAY
2. AIR PLUG
3. ARM CUFF
4. AIR JACK
5. CUFF TUBE
6. START/STOP BUTTON
7. RECALL MEMORY BUTTON
8. MODE BUTTON
9. SET BUTTON
10. UM-4(AA)1.5 V x 4 
 ALKALINE BATTERIES

*Caution!
Substitution of a component 

different from that supplied 

might result in measurement 
error.

* Figure No.5

5. Attaching arm cuff

5-1. Press your brachial artery by two fingers approximately 1 inch 
 above the elbow on the inside of your left arm.
5-2. Determine where your strongest pulse is. Slide the end of arm 
 cuff furthest from the tube through the metal ring to a loop.
 The smooth cloth should be on the inside of the cuff.
5-3. If the cuff is located correctly, the Velcro will be on the outside 
 of the cuff and metal ring will not touch your skin.
5-4. Put left arm through the cuff loop. The bottom of the cuff should 
 be approx. 0.5 inch above elbow. The cuff should lie over the 
 brachial artery on the inside of the arm.
5-5. Pull the cuff so that the top and bottom edges are tightened 
 around your arm.
5-6. When the cuff is positioned properly, press the Velcro firmly 
 against the pile side of the cuff.
5-7. Sit on a chair and place your arm on the table so that the cuff is 
 at the same level as your heart.
5-8. Relax your arm and turn your arm upward.
5-9. Make sure there are no kinks in the air tube.

Elbow

Pile Side
Material

Smooth

Velcro

Metal ring

Bottom Edge

1/2 Inch(1.5cm)

Cuff Tube

Cuff Tab

6-1. Insert batteries in position:

6. How to Measure

A. Open battery cover.
B. Refer to the picture in positioning batteries.
C. All segments will appear on the display in 3 seconds.

6-2. Setting Date and Time:

A. Press "MODE" button ("month" starts flashing). Press "SET" button to set the 
 current month.
B. Press "MODE" button ("date" starts flashing). Press "SET" button to set current 
 date.
C. Press "MODE" button ("hour" starts flashing). Press "SET" button to set current 
 hour.
D. Press "MODE" button ("minute" starts flashing). Press "SET" button to set current
 minute.
E. Press "MODE" button to save the settings.

6-3. Steps to take blood pressure measurement:

A. Press "SET" button to select a particular memory zone (1,2, or 3).
B. Press "START/STOP" button to start 
 measurement.
 The monitor will automatically inflate to 
 the level that is right for you.

C. After inflation of the cuff, the pressure will slowly decrease and when a pulse is
 detected, the HEART RATE SYMBOL will start flashing.

D. When the measurement is finished, the systolic pressure, diastolic pressure, 
 heart rate will be displayed for 1 minute.
E. The monitor will automatically shut off after 1 minute if without any operation.

c. Press "MEMORY" button. The latest measurement data will appear with the 
 number as well as the recording date and time. Data are recalled from the 
 latest measurement to the oldest.
d. Press "MEMORY" button to read the next measurement data.
e. To stop reading stored data, press "START/STOP" button.

6-4. Storing, recalling and erasing measurement data:
A. Storing data:

After each measurement, the systolic and diastolic pressure, heart rate with the 
time and date will be automatically stored. The monitor features a 3 zone memory 
capability. Each memory zone holds the last 40 measurements, replacing the 
oldest data with new one.

B. Recalling data:

a. Press "SET" button to select memory zone.
b. Press "MEMORY" button to enter Memory Mode. The average of last 3 
 measurement data will first appear on the display.

C. Erasing data:

a. Press "SET" button to select the memory zone (1, 2 or 3), then press
 "MEMORY" button.
b. Press and hold "MODE" and "SET" buttons simultaneously, all stored data,
 including the average reading, in the selected memory zone will be erased.
c. To confirm the data has been erased, press "MEMORY" button, no data 
 (except month, date, time and memory zone number) should appear.

*

te!

 If there is no data in the selected memory zone, nothing (except month, date, 

 time and memory zone) will appear on the display.

 If there is 1 data in the selected memory zone , the measurement data will 

 appear on the display.

 If there are 2 data in the selected memory zone, the average of the 2 data will 

 appear on the display.

*

te!

When taking repeat measurements, make sure to select the same memory
zone so that the measurements are recorded in the appropriate memory.

*

te!

If the cuff does not stop inflating, remove the cuff at once.

10

 

 

 

 

 

 

 

 

 

Display mark

Corrective action

Battery symbol

Measurements in progress.
Do not move or talk.

Replace all four batteries with new ones. 
Insert the batteries in correct polarities.

Pulse 

symbol

Flashes on the LCD when pulse is 
detected during measurement.

Condition/Cause

Appears when the battery voltage is
excessively low.

Measuring

error symbol

Press "START/STOP" button again to re-measure. 
Or check cuff if wrapped properly around the arm 
and according to instructions. Check palm if exerting 
effort. Check if there is movement during 
measurement. Check if posture is correct.

Appears when the accurate blood 
pressure could not be obtained.

8. Display Messages/Troubleshooting

 To ensure precise measurement 

reading, recalibration of the device is 
recommended after 2 years from the 
date of purchase. Shipping plus 
handling cost and recalibration service 
fee shall be charged accordingly.

P/N: 323101291 VER:001

TIME

M

SYS.

mmHg

DIA.

mmHg

PULS

E

/MIN

.

DATE

Sponsored links

Jump to page: 1

 

GO

< Prev

Next >

Page 1

Instruction Manual UPPER ARM BLOOD PRESSURE MONITOR 

 Do not wrap the cuff over jacket / sweater sleeve, or measurement cannot be done.

 Keep in mind that blood pressure naturally varies from time to time throughout the day and is affected by lots of different factors such as smoking,