CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-122
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.13.11 Generator Configuration Backup and Restore
NOTE:
A PERMANENT BACKUP FILE OF THE GENERATOR CONFIGURATION (RECEPTOR SETUP, AEC SETUP AND
CALIBRATION, ETC,) SHOULD BE KEPT IN THE EVENT THE GENERATOR SOFTWARE IS CORRUPTED OR A “FACTORY
DEFAULTS” IS REQUIRED. THE BACKUP FILE SHOULD BE STORED ON AN EXTERNAL STORAGE DEVICE, SUCH AS A
LAPTOP OR USB FLASH DRIVE.
NOTE:
ENSURE THERE IS AN EXTERNAL USB DEVICE CONNECTED TO THE TOUCHSCREEN CONSOLE USING ONE OF THE TWO
AVAILABLE USBA PORTS AT THE REAR OF THE TOUCHSCREEN CONSOLE.
Figure 3C-46: Generator Configuration Backup and Restore utility
Step
Action
1.
Touchscreen GenWare
®
must be launched before proceeding. This is done from the
GenWare
®
button on the touchscreen
System Utilities
menu.
2.
Press the
button to launch the
Backup and Restore
utility.