Dry Heat Sterilizer models 200 and 300
Operating Instructions
MODEL
200
MODEL 300
2364 Leicester Road, P.O. Box 175, Leicester, NY 14481
Phone (585) 382-3223 ● Fax (585) 382-9481 ●
www.cpac.com
Rev 4, 18 October, 2019
Страница 1: ...ry Heat Sterilizer models 200 and 300 Operating Instructions MODEL 200 MODEL 300 2364 Leicester Road P O Box 175 Leicester NY 14481 Phone 585 382 3223 Fax 585 382 9481 www cpac com Rev 4 18 October 20...
Страница 2: ...eter Thermostat 6 Pilot Lights 6 Accessories 6 Chemical Indicator Use 6 Spore Test Strips 7 Cleaning and Maintenance 7 Power Source and Ratings 7 Simulated Use Testing 7 Environmental conditions 9 Ele...
Страница 3: ...ead and understand all instructions 2 Care must be taken to avoid burns from touching hot parts 3 Do not operate this appliance with a damaged cord or if appliance has been dropped or damaged until it...
Страница 4: ...ears by dentists hospitals and the U S Armed Forces This design has been refined and improved over the years to make it safe and simple to operate easy to use and low in cost to operate It is user fri...
Страница 5: ...5 Steri DENT Part No 400636 NYLON SELF SEAL POUCHES 2 X 10 Steri DENT Part No 400651 NYLON SELF SEAL POUCHES 3 X 10 Steri DENT Part No 400637 NYLON SELF SEAL POUCHES 4 X 10 Steri DENT Part No 400638 N...
Страница 6: ...ents use only dry heat packaging material suitable for 392 F 200 C temperatures All items being sterilized and any associated wraps should be able to withstand temperatures up to 200 C 392 F Instrumen...
Страница 7: ...the unit has power The right light is on only while the heating elements are operating and heating the sterilization chamber Once the chamber has reached sterilization temperature the light will come...
Страница 8: ...ts that require no maintenance An authorized service representative or CPAC should be contacted if any of these parts need to be replaced Replace the inner door gasket once a year or sooner if it beco...
Страница 9: ...8 Figure 2 Test Load Configuration for Model 2100 Figure 3 Test Load Configuration for Model 3100...
Страница 10: ...easing linearly to 50 at 40 C 104 F Pollution Degree 2 applies in accordance with IE 664 Transient Over voltage Category II applies Supply voltage not to fluctuate more than 10 12V at 120V 22V at 220V...
Страница 11: ...Was cycle observation data recorded temperature readings at 1st and 5 minute intervals 6 Refer to Biological Testing Protocols section for further guidance Thermometer not reading 320 330 F Verify th...
Страница 12: ...the ON position If not lit when in the HOLD position or during any sterilization cycle replace the light 3 The right light illuminates only while the heating elements are operating and heating the st...
Страница 13: ...re labeled for use in a dry heat sterilizer 3 Bags pouches are not left in sterilizer for more than 15 minutes after cycle is done Adjust thermostat to obtain correct thermometer range Replace thermos...
Страница 14: ...ge Replace circuit switch Replace fuse fuse holder Replace bad element Timer not timing down 1 Timer turned past 10 minutes to test function 2 Empty sterilizer and turn timer to 120 minute setting to...
Страница 15: ...20 x 16 x 14 Model 300 45 lbs 22 x 18 x 18 Tray size Model 200 2 trays 10 W x 6 25 D 254 mm x 159 mm Model 300 3 trays 11 75 W x 7 25 D 298 mm x 184 mm Electrical 110V 120V 60 Hz or 220V 240V 50 Hz _...
Страница 16: ...ra Tray for Model 300 12 x 7 301095 Door Gasket Kit SD 300 301108 Side Heating Elements SD 300 301058 Bottom Heating Element SD 300 301059 COMMON PARTS BOTH MODELS Rubber Foot CX0051 Tray Handle Both...
Страница 17: ...of the tray NOTE Model 200 tray weighs 0 3 lbs Model 300 tray weighs 0 4 lbs Self Sealing nylon pouches from Steri DENT are recommended Steri DENT Part No 400636 NYLON SELF SEAL POUCHES 2 X 10 Steri...
Страница 18: ...days are minimally required to obtain results from a biological indicator test Note There are no rapid read biological indicators currently for dry heat sterilization as there are for steam and ethyl...
Страница 19: ...ayer instruments in the instrument tray with no instrument overlap 3 Add a chemical indicator to each pouch or place chemical indicator strip under an instrument in each tray to secure it in place 4 I...
Страница 20: ...p during shipment b If conducting in office testing of the spore strip use sterile techniques when removing the spore strip from its envelope and transferring the strip to the media tube for incubatio...
Страница 21: ...that CDC states that the large margin of safety required for sterilization technologies documented 12 Log spore kill that there is minimal infection risk associated with items in a load that show spor...
Страница 22: ...call CPAC Equipment 800 828 6011 and ask for a service technician to discuss the problem and to determine a cause for failure Provide the technician with information necessary for determination of fa...
Страница 23: ...inutes intervals for the duration of the cycle Sterilizer readings recorded in the test data manual Sterilizer readings throughout the cycle were between 320 330 F Post Check After Completion of Steri...
Страница 24: ...damages to the carrier This warranty gives you specific legal rights You may also have additional rights that vary from state to state Any unit to be repaired under warranty must be shipped freight p...