Scanning Electron Microscope
EM-30PLUS Operating Manual
54
4
Preparation for Taking Photos
4.1 Power ON/OFF
4.1.1 Turning on the Main Body of EM-30PLUS
1) Turn on the main power on the back of EM-30PLUS
product.
- Keep the power of equipment on.
2) Turn on the power switch on the side of EM-
30PLUS.
(when power is on, the switch turns red)
3) The rotary pump on the back of the product starts
to work.
4) When operation of rotary pump finishes, turbo pump starts to work and makes
chamber inside vacuum state.
Rotary pump is in air pressure 10
-1
torr: vacuum state reaches 50%
Turbo molecular pump is in vacuum state within a range of 10
-1
torr~10
-4
torr: it
reaches 100% of vacuum state.
In general it takes about 2.5 minutes until it reaches 100% of vacuum in air pressure.
However, the time for 100% vacuum setting varies depending on specimens and
operation condition of equipment.
5) When you double click the execution icon on the main
screen of control PC, NS is executed.