Error Codes
B-2
Service Manual
B.2
Error Codes and Suggested Resolutions
Error Categories:
•
A/D —
Errors associated with or reported by the A/D microprocessor.
•
POST —
Errors occurring during power-on self-test.
•
UI —
Errors associated with or reported by the UI microprocessor.
Note:
For most errors, the initial corrective action is to power-cycle the warming unit. If the
error persists, continue with the recommendations listed for the error.
003
Category:
A/D
Alarm Priority:
Low
Description:
Checksum error on code in flash. A/D flash is defective.
Resolution:
Try reloading software to the unit. See
If this does not resolve the issue, replace the power PCBA. See
004
Category:
A/D
Alarm Priority:
Low
Description:
One of the following conditions occurred:
•
Temperature was more than ±1°C from the setpoint for at least 30 consecu-
tive seconds after the setpoint was reached.
•
Temperature did not reach the specified setpoint within 10 minutes.
Resolution:
Check for:
•
Blockages in the air intake, hose, and filter.
This error may also appear due to issues at the customer site:
•
Low mains voltage
•
Ambient temperature outside of specification.
Содержание WarmTouch
Страница 1: ...Service Manual WarmTouch Convective Warming Unit...
Страница 8: ...Page Left Intentionally Blank vi Service Manual...
Страница 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Страница 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Страница 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Страница 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Страница 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Страница 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Страница 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Страница 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Страница 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Страница 176: ...Page Left Intentionally Blank 10 58 Service Manual Internal Component Replacement Inside Front Enclosure...
Страница 232: ...Page Left Intentionally Blank 11 56 Service Manual Internal Component Replacement Inside Rear Enclosure...
Страница 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Страница 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Страница 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Страница 313: ......