System Setup
Valleylab FT10 Energy Platform User’s Guide
3-11
Service and Settings
Service Menu
Items in the Covidien Service menu are specifically for qualified service personnel and are
password protected. The password is 423213. The Service Menu button accesses settings
for
•
RF Output Test
•
Touchscreen Calibration
•
Energy Calibration
•
Country Setup
•
RFID Antenna Control
RF Output Test
The
RF Output Test Screen
is used to test the RF output of the generator. It allows a
technician to output RF in a specified mode, to a specified energy channel, at a specified
power level. While RF is output, the
RF Output Test Screen
displays the internal sensor
readings of average power, RMS voltage, RMS current, impedance magnitude, impedance
phase angle, REM impedance magnitude, and REM impedance phase angle.
Touchscreen Calibration
The Touchscreen Calibration screen allows the technician to calibrate the touchscreen. See
the
Valleylab FT10 Energy Platform Service Manual
for more information.
Energy Calibration
The Energy Calibration screen allows the technician to calibrate the RF output. See the
Valleylab FT10 Energy Platform Service Manual
for more information.
Country Setup
The Country Setup screen allows the technician to identify in which country the generator
is set up to operate. If this is the first time using the generator, the country code needs to
be selected before the LigaSure/Bipolar energy channel will function. See the
First Time
Setup Guide
for instructions on how to set the country code.
RFID Antenna Control
The RFID Antenna Control screen allows the technician to manually initiate an RFID scan
of an instrument connected to the LigaSure receptacle.