SurgiStat Low-Power Electrosurgical Generator Service Manual
Safety Precautions When Testing the Generator
Fire/Explosion Hazards
Precaution
Accessories must be connected to the proper receptacle type. In particular, bipolar accessories must
be connected only to the bipolar instrument receptacle. Improper connection may result in
inadvertent generator activation or a contact quality monitor alarm.
Set power levels to the lowest setting before testing an accessory.
Notice
During bipolar electrosurgery, do not activate the generator until the forceps have made contact
with the patient. Product damage may occur.
Warning
Explosion Hazard
Do not use the generator in the presence of flammable anesthetics, gases,
liquids, or objects.
Fire Hazard
Do not place active accessories near or in contact with flammable materials (such as
gauze or surgical drapes). Electrosurgical accessories that are activated or hot from use can cause
a fire. Use a holster to hold electrosurgical accessories safely away from personnel and flammable
materials.
Fire/Explosion Hazard—
The following substances will contribute to increased fire and explosion
hazards in the operating room:
•
Flammable substances (such as alcohol based skin prepping agents and tinctures)
•
Naturally occurring flammable gases which may accumulate in body cavities such as the bowel
•
Oxygen-enriched atmospheres
•
Oxidizing agents (such as nitrous oxide [N
2
0] atmospheres).
The sparking and heating associated with electrosurgery can provide an ignition source. Observe
fire precautions at all times. When using electrosurgery in the same room with any of these
substances or gases, prevent their accumulation or pooling under surgical drapes, or within the
area where electrosurgery is performed.
The use of flammable anesthetics or oxidizing gases such as nitrous oxide (N
2
O) and oxygen should
be avoided if a surgical procedure is carried out in the region of the thorax or the head, unless these
agents are evacuated.
Non-flammable agents should be used for cleaning and disinfection wherever possible.
Содержание SurgiStat
Страница 1: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator ...
Страница 2: ......
Страница 3: ...Service Manual SurgiStat TM Low Power Electrosurgical Generator 1079008 ...
Страница 10: ......
Страница 34: ......
Страница 54: ......
Страница 74: ......
Страница 88: ......
Страница 100: ...9 12 SurgiStat Low Power Electrosurgical Generator Service Manual Connector Panel Block Connector Panel Block ...
Страница 110: ......
Страница 111: ......