Figure 2-61. Air and oxygen flow during retraction
High oxygen flows
Piston retraction speed (L/min)
High-flow solenoid on/off
Oxygen flow from high-flow solenoid (L/min)
Air flow from air intake manifold (L/min)
Low oxygen flows
Piston retraction speed (L/min)
Low-flow solenoid on/off
Oxygen flow from low-flow solenoid (L/min)
Air flow from air intake manifold (L/min)
100% room air flows
Piston retraction speed (L/min)
Air flow from air intake manifold (L/min)
150 L/min
150 L/min
150 L/min
15 L/min
On
Off
Off
On
retraction
135 L/min
150 L/min
150 L/min
150 L/min
150 L/min
7-00169
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......