Alarms and Troubleshooting
3-2
Puritan Bennett 540™ Ventilator User’s Manual
3.2
Alarm Display
During operation, when an alarm is activated:
•
One of the red or yellow alarm indicators to the left of the
ALARM CONTROL
key
illuminates and flashes.
•
An alarm tone sounds.
•
A message is displayed and flashes in reverse video at the bottom of the Ventilation Menu or
Alarm Menu.
Figure 3-1.
Alarm Displays
Note:
There are currently no Low Priority (LP) Alarms.
When an alarm is triggered, if the current menu displayed is not the Ventilation parameters or Alarm
menu, the display automatically switches to one of these menus to display the alarm message.
Note:
If there is no corrective action and if the audible alarm is not inhibited (Audio Paused) or reset (Alarm Reset)
within 60 seconds, High Priority alarms will sound at the maximum level of 85 dB(A).
V
ENTILATION
M
ENU
A
LARM
M
ENU
A
LARM
I
NDICATORS
A
LARM
M
ESSAGES
Содержание Puritan Bennett 540
Страница 1: ...Puritan Bennett 540 Ventilator User s Manual...
Страница 6: ...This page intentionally blank...
Страница 8: ...This page intentionally blank...
Страница 10: ...This page intentionally blank...
Страница 80: ...Installation and Assembly 4 20 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 86: ...Operating Procedures 5 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 94: ...This page intentionally blank...
Страница 100: ...Routine Maintenance 8 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Страница 124: ...Unpacking and Preparation C 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 140: ......