Description of Functional Subsystems
Service Manual
A-11
A.6.1
Inhalation System
The inhalation system supplies gas according to the user-selected breath delivery settings.
Figure A-6.
Newport™ e360 Ventilator Inhalation System
Supply gas enters the inlet filters (one for oxygen and one for air) at a pressure of 30 to 90 pounds
per square inch gauge (psig). Each inlet filter traps particles as small as 40 microns and includes a
water trap to collect any water from the supply gas.
The filtered supply gas then enters a regulator that regulates pressure from 30 to 90 psig down to
15 psi. Transducers monitor pressure at the regulators, and a pressure drop at either regulator
generates a low inlet pressure alarm.
1
Air inlet
9
Inhalation block
2
Crossover solenoid
10
Safety regulator
3
PSOL valve air
11
Safety solenoid
4
Air regulated pressure
12
Pressure oxygen regulated
5
Exhalation regulator
13
Inlet block oxygen
6
Air
14
Oxygen inlet
7
Machine pressure 1
sensor
15
Inlet block air
8
Rezero valve machine 1
16
Disk filters
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......