6-1
6 Operational Verification
6.1
Overview
The operational verification procedure ensures that the Newport™ e360 ventilator is in proper
operating condition.
WARNING:
Do not use the ventilator unless it passes the operational verification procedure.
6.2
Test Equipment
•
Electrical safety test (EST) analyzer
•
Certified calibration analyzer manufactured by TSI, Inc., or equivalent
•
Oxygen analyzer
•
Reusable patient breathing circuit (PBC340A), or equivalent
•
Adult test lung (LNG800P) or adult test lung with RP20 airway resistor (LNG600A), or equivalent
•
Adjustable (0-60 psig) dry, clean and regulated medical-grade air and oxygen gas supplies
•
A cap to occlude the patient circuit Y-piece (CAP100P), or equivalent
Caution:
The accuracy of all test equipment (electronic or pneumatic) used for verification or calibration
procedures should be certified annually by a testing laboratory before use.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......