Factory Defaults
Operator’s Manual
43
4.5.2
Visual Alarm Indicators
Visual alarms appear on the screen in order of highest priority, regardless of
any audible alarm status. Reference
4.6
Factory Defaults
The monitoring system ships with factory default settings. To set different insti-
tutional default settings, contact a qualified service technician.
Table 4-4.
Parameter Ranges and Factory Defaults
Parameter
Ranges/Selection
Factory Default
Adult
Pediatric
Neonatal
SpO
2
%SpO
2
Upper Alarm Limit
21 to 100% (1% steps)
100%
95%
%SpO
2
Lower Alarm Limit
20 to 99% (1% steps)
90%
85%
%SpO
2
Limit Alarm Inhibition
On, Off
Off
SatSeconds™ Alarm
Off, 10, 25, 50, 100
100
Pulse Rate
Pulse Rate Upper Alarm Limit
30 to 245 bpm (5 bpm steps)
170 bpm
200 bpm
Pulse Rate Lower Alarm Limit
25 to 240 bpm (5 bpm steps)
50 bpm
75 bpm
100 bpm
PR Limit Alarm Inhibition
On, Off
Off
Tabular Trends
Scroll
1, 5, 100, 500
1
Graphical Trends
SpO
2
On, Off
On
PR
On, Off
On
Others
Patient Mode
Adult, Pediatric, Neonatal
Adult
Alarm Volume
1, 2, 3, 4, 5, 6, 7, 8
5
Key Beep Volume
Off, 1, 2, 3, 4, 5, 6, 7
4
Pulse Volume
Off, 1, 2, 3, 4, 5, 6, 7
4
Содержание Nellcor
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Страница 8: ...viii ...
Страница 10: ...Page Left Intentionally Blank x ...
Страница 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Страница 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...
Страница 119: ......