Sound Pressure
Operator’s Manual
95
11.7
Sound Pressure
Table 11-4.
Pulse Oximetry Sensor Accuracy and Ranges
Range Type
Range Values
Measurement Ranges
SpO
2
saturation range
1% to 100%
Pulse rate range
20 to 250 beats per minute (bpm)
Perfusion range
0.03% to 20%
Display sweep speed
6.25 mm/sec, 12.5 mm/sec, 25.0 mm/sec
Measurement Accuracy
Pulse rate accuracy
20 to 250 beats per minute (bpm) ±3 digits
SpO
2
saturation accuracy
1
1. Monitoring system measurements are statistically distributed; about two-thirds of
monitoring system measurements can be expected to fall in this accuracy (ARMS) range.
Reference the Clinical Studies section for test results. For a complete listing of SpO2 accuracy
across the full line of available Nellcor™ sensors, contact Covidien, a local Covidien
representative, or locate it online at www.covidien.com.
70% to 100% ±2 digits, neonates: ±3 digits
Operating Range and Dissipation
Red Light Wavelength
Approximately 660 nm
Infrared Light Wavelength
Approximately 900 nm
Optical Output Power
Less than 15 mW
Power Dissipation
52.5 mW
Table 11-5.
Sound Pressure in Decibels
Alarm Type
Volume Setting
High (7-8)
Med High (5-6)
Med Low (3-4)
Low (1-2)
High Priority
83.6-87.4 dB
74.1-77.9 dB
65.6-69.5 dB
57.6-61.1 dB
Medium Priority
82.0-84.7 dB
70.2-74.8 dB
64.5-66.9 dB
53.6-57.9 dB
Low Priority
77.2-81.7dB
69.5-72.6 dB
60.1-63.8 dB
50.8-56.0 dB
Содержание Nellcor
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Страница 8: ...viii ...
Страница 10: ...Page Left Intentionally Blank x ...
Страница 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Страница 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...
Страница 119: ......