Product Specifications
122
Operator’s Manual
Rating of Nurse Call Relay
Maximum Input Voltage
30 VAC or VDC (polarity is not important)
Load Current
120 mA continuous (peak 300 mA @ 100 ms)
Minimum Resistance
26.5 ohms to 50.5 ohms (40.5 ohms typical) during alarms
Ground Reference
Isolated Ground
Electrical Isolation
1500 Volts
Environmental Conditions
Operating
Temperature
5 ºC to 40 ºC (41 ºF to 104 ºF)
Altitude
-390 m to 3,012 m
(-1,254 ft. to 9,882 ft.)
Atmospheric Pressure
70 kPa to 106 kPa
(20.6 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non-condensing
Transport and Storage
Not in shipping container
In shipping container
Temperature
-20 ºC to 60 ºC
-20 ºC to 70 ºC
(-4 ºF to 140 ºF)
(-4 ºF to 158 ºF)
Altitude
-390 m to 5,574 m
-390 m to 5,574 m
(-1,254 ft. to 18,288 ft.)
(-1,254 ft. to 18,288 ft.)
Atmospheric Pressure
50 kPa to 106 kPa
50 kPa to 106 kPa
(14.7 in. Hg to 31.3 in. Hg)
(14.7 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non-condensing 15% to 95% non-condensing
Performance Specifications
Measurement Range
SpO
2
Saturation Range
1% to 100%
Pulse Rate Range
20 to 250 beats per minute (bpm)
Perfusion Range
0.03% to 20%
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......