Technical Discussion
106
Automatic Calibration
Because light absorption by hemoglobin is wavelength dependent and
because the mean wavelength of LEDs varies, an oximeter must know the
mean wavelength of the sensor's red LED to accurately measure SpO
2
. During
manufacturing, the mean wavelength of the red LED is encoded in the sensor.
During monitoring, the N-560's software uses the coefficients that are
appropriate for the wavelength of that individual sensor's red LED; these
coefficients are then used to determine SpO
2
. The N-560’s software obtains
coefficients when the N-560 is turned on, periodically thereafter, and each
time a new sensor is connected.
Additionally, to compensate for differences in tissue thickness, the light
intensity of the sensor's LEDs is adjusted automatically.
Functional Versus Fractional Saturation
This N-560 measures functional saturation -- oxygenated hemoglobin
expressed as a percentage of the hemoglobin that can transport oxygen. It
does not detect significant amounts of dysfunctional hemoglobin, such as
carboxyhemoglobin or methemoglobin. In contrast, hemoximeters such as the
IL482 report fractional saturation -- oxygenated hemoglobin expressed as a
percentage of all measured hemoglobin, including measured dysfunctional
hemoglobins. To compare functional saturation measurements to those from
an instrument that measures fractional saturation, fractional measurements
must be converted as follows:
Measured versus Calculated Saturation
When saturation is calculated from a blood gas partial pressure of oxygen
(PO
2
), the calculated value may differ from the SpO
2
measurement of a pulse
oximeter. This usually occurs because the calculated saturation was not
appropriately corrected for the effects of variables that shift the relationship
between PO
2
and pH, temperature, the partial pressure of carbon dioxide
(PCO
2
), 2,3-DPG, and fetal hemoglobin. See Figure 46.
functional
saturation
=
fractional
saturation
100
-
(
%
carboxyhemoglobin
+
%
methemoglobin
)
x
100
Содержание Nellcor OxiMax N-560
Страница 1: ...Service Manual OxiMaxTM Technology Nellcor OxiMax N 560 Pulse Oximeter TM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 127: ...Technical Discussion N 560 Service Manual 117 Figure 49 Main PCB Battery Charger Schematic Diagram...
Страница 128: ...Technical Discussion 118...
Страница 129: ...Technical Discussion N 560 Service Manual 119 Figure 50 Main PCB Isolation for NELL 3 Schematic Diagram...
Страница 130: ...Technical Discussion 120...
Страница 131: ...Technical Discussion N 560 Service Manual 121 Figure 51 Main PCB Isolation for External Interface Schematic Diagram...
Страница 132: ...Technical Discussion 122...
Страница 133: ...Technical Discussion N 560 Service Manual 123 Figure 52 Main PCB DC Battery Input and DC DC Converter Schematic Diagram...
Страница 134: ...Technical Discussion 124...
Страница 135: ...Technical Discussion N 560 Service Manual 125 Figure 53 Main PCB Interface and Speaker Control Schematic Diagram...
Страница 136: ...Technical Discussion 126...
Страница 137: ...Technical Discussion N 560 Service Manual 127 Figure 54 Main PCB MCU and Storage Control Schematic Diagram...
Страница 138: ...Technical Discussion 128...
Страница 139: ...Technical Discussion N 560 Service Manual 129 Figure 55 Main PCB Parts Locator Diagram...
Страница 140: ...Technical Discussion 130...
Страница 141: ...Technical Discussion N 560 Service Manual 131 Figure 56 Front Panel PCB 7 Segment Display Schematic Diagram...
Страница 142: ...Technical Discussion 132...
Страница 143: ...Technical Discussion N 560 Service Manual 133 Figure 57 Front Panel PCB LED Display and Buttons Schematic Diagram...
Страница 144: ...Technical Discussion 134...
Страница 145: ...Technical Discussion N 560 Service Manual 135 Figure 58 Front Panel PCB Interface and LED Drivers Schematic Diagram...
Страница 146: ...Technical Discussion 136...
Страница 147: ...Technical Discussion N 560 Service Manual 137 Figure 59 Front Panel Parts Locator Diagram...
Страница 148: ...Technical Discussion 138...
Страница 149: ...Technical Discussion N 560 Service Manual 139 Figure 60 Power Supply Schematic Diagram...
Страница 150: ...Technical Discussion 140...
Страница 151: ...Technical Discussion N 560 Service Manual 141 Figure 61 Power Supply Parts Locator Diagram...
Страница 152: ...Technical Discussion 142...
Страница 156: ......
Страница 157: ......