Service Mode
Service Manual
6-19
3.
Turn on the monitoring system by pressing the POWER ON key.
4.
Touch the monitoring screen at the prompt to enter Service Mode.
Note:
The monitoring system will eventually continue to boot if the user does not touch
the monitoring screen at the Service Mode prompt.
Figure 6-10.
Prompt to Enter SERVICE MODE
5.
Enter 62907 as the service mode password at the prompt, using the onscreen
number pad.
6.
Press ENTER SERVICE MODE.
7.
Select the desired menu option and any associated submenu options.
6.5.1
Settings Menu
Import and Export Settings
Both import settings and export settings function in the same fashion. Both
require a USB flash drive. Reference
To export master institutional defaults
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......