Performance Tests
Service Manual
10-27
5.
Connect the SRC-MAX tester to the interface cable.
6.
Verify the listed system behaviors.
Figure 10-6.
BPM Test: BPM 60 and SpO
2
75
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive, creating an SpO
2
LOW
alarm condition.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately 1/2-inch peak to peak (P-T-P) amplitude.
Actual amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
7.
Press PULSE RATE.
8.
Verify the 200 BPM LED lights. The monitoring system registers pulse rate increas-
es and stabilizes to a value between 194 to 206 BPM inclusive.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......