Performance Tests
Service Manual
10-19
14.
Press ALARM SILENCE.
15.
With the alarm silenced, verify the following:
a.
Alarm remains silenced for two (2) minutes.
b.
Alarm Silence indicator lights and time remaining counts down
c.
SpO
2
and BPM alarms continue to flash yellow
d.
SpO
2
and BPM values are surrounded by yellow
e.
Pulse tone is audible
f.
Audible alarm returns in approximately two (2) minutes
Figure 10-3.
Sample: Configuring Alarms with Silenced Alarm
To configure alarm volume
Configure the desired volume level of the alarm, from a silent 0% to a loud
100%. The factory default volume level is set to 75%.
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Select SOUND SETTINGS.
4.
Select ALARM VOLUME.The monitoring system displays an volume icon with
increment and decrement buttons.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......